Focused Ultrasound Sonications of Tumor Model in Head Phantom under MRI Monitoring: Effect of Skull Obstruction on Focal Heating
Journal of Medical Physics · 2025-01 · 50 (1) · 38-45
DOI: 10.4103/jmp.jmp_177_24 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Physics, online erschienen am 07.03.2025. Verfasst von Anastasia Antoniou und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).
Abstract (Originalsprache)
Purpose
This study presents the outcomes of a series of magnetic resonance imaging (MRI)-guided focused ultrasound (MRgFUS) sonications performed on an anatomically accurate head phantom with an embedded tumor simulator to evaluate the effectiveness of partial and complete tumor ablation with obstruction from thin polymer skull mimics. Materials and Methods: The tumor simulator was subjected to single and grid sonications using a single-element concave transducer integrated with an MRI-compatible focused ultrasound (FUS) robotic system. All experiments were carried out in a high-field MRI scanner utilizing proton resonance frequency thermometry and T2-weighted (T2-W) turbo spin echo (TSE) imaging to evaluate the induced thermal effects. FUS transmission through 1-mm thick three-dimensional-printed polymer skull mimics was compared to unobstructed sonication through a circular aperture in the skull model.
Results
T2-W TSE imaging demonstrated sharp contrast between the tumor and hyperintense FUS lesions. Complete tumor coverage was achieved through robotic-assisted grid ablation without a skull mimic, as well as with a 1-mm resin skull mimic intervening in the beam. With the lowest attenuation among tested polymers, the resin skull resulted in approximately a 20% reduction in focal temperature change compared to unobstructed sonication, yet still facilitated sharp beam focusing, raising the tumor temperature to ablative levels.
Conclusions
The study provides preliminary evidence for the potential application of a thin biocompatible implant to temporarily replace a skull portion facilitating MRgFUS ablation of inoperable tumors using a single-element transducer. The tumor-embedded head phantom was proven effective for testing MRgFUS oncological protocols and equipment.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
- Department of Diagnostic and Interventional Radiology, German Medical Institute, Limassol, Cyprus.
- Department of Radiation Oncology, German Medical Institute, Limassol, Cyprus.
- Department of Veterinary Medicine, University of Nicosia School of Veterinary Medicine, Nicosia, Cyprus.
- VET EX MACHINA Limited, Nicosia, Cyprus.
- Department of Electronics and Information Engineering, Hangzhou Dianzi University, Hangzhou, Zhejiang, China.
Zitieren
APA
Antoniou, A., Chrysanthou, A., Georgiou, L., Christofi, A., Roussakis, Y., Ioannides, C., Spanoudes, K., Zhao, J., Yu, L., & Damianou, C. (2025). Focused Ultrasound Sonications of Tumor Model in Head Phantom under MRI Monitoring: Effect of Skull Obstruction on Focal Heating. Journal of Medical Physics, 50(1), 38-45. https://doi.org/10.4103/jmp.jmp_177_24
Vancouver
Antoniou A, Chrysanthou A, Georgiou L, Christofi A, Roussakis Y, Ioannides C, et al. Focused Ultrasound Sonications of Tumor Model in Head Phantom under MRI Monitoring: Effect of Skull Obstruction on Focal Heating. Journal of Medical Physics. 2025;50(1):38-45. doi:10.4103/jmp.jmp_177_24
BibTeX
@article{Antoniou2025,
author = {Antoniou, Anastasia and Chrysanthou, Antreas and Georgiou, Leonidas and Christofi, Antonis and Roussakis, Yiannis and Ioannides, Cleanthis and Spanoudes, Kyriakos and Zhao, Jufeng and Yu, Liyang and Damianou, Christakis},
title = {Focused Ultrasound Sonications of Tumor Model in Head Phantom under MRI Monitoring: Effect of Skull Obstruction on Focal Heating},
journal = {Journal of Medical Physics},
year = {2025},
volume = {50},
number = {1},
pages = {38--45},
doi = {10.4103/jmp.jmp_177_24}
}
EUVIMED