Effects of skull properties on continuous-wave transcranial focused ultrasound transmission

Han Li iD, Isla Barnard, Tyler Halliwell, Xinyu Zhang, Andreas Melzer, Zhihong Huang

The Journal of the Acoustical Society of America · 01.04.2025 · 157 (4) · 2336-2349

DOI: 10.1121/10.0036344 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of the Acoustical Society of America, online erschienen am 02.04.2025. Verfasst von Han Li und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Science and Engineering, University of Dundee, Dundee, DD1 4HN, United Kingdom. Hauptthemen laut MEDLINE: Skull, Ultrasonic Therapy. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 52 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Transcranial low-intensity focused ultrasound can deliver energy to the brain in a minimally invasive manner for neuromodulation applications. However, continuous sonication through the skull introduces significant wave interactions, complicating precise energy delivery to the target. This study presents a comprehensive examination of intracranial acoustic fields generated by focused ultrasound transducers and assesses the characteristics of cranial bone that affect acoustic transmission. Acoustic field maps were generated at 88 regions of interest across 10 historical and 2 Thiel-embalmed human skull specimens with sonication at frequencies of 220, 650, and 1000 kHz. The average peak pressure insertion losses for historical skulls were 3.6 ± 3.4, 9.3 ± 3.3, and 14.8 ± 5.8 dB, respectively, and for Thiel skulls, the respective losses were 2.9 ± 1.8, 9.4 ± 2.6, and 17.0 ± 5.5 dB. The effects of skull thickness, skull density ratio, and skull curvature on intracranial peak pressure, power, and focal area were investigated and linear fits produced. Several unfavorable focusing performances were observed in regions with excessive thickness variation. The effects of angulation and spacing between the transducer and the skull were also investigated. Preliminary findings indicate that wave superposition resulting from skull and transducer spacing could lead to a 30%–40% uncertainty in peak recorded intracranial pressure.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SkullUltrasonic TherapyHumansIntracranial PressureUltrasonic WavesTransducersCadaverMale

Autor:innen und Institutionen

  • School of Science and Engineering, University of Dundee, Dundee, DD1 4HN, United Kingdom.
  • School of Physics, Engineering and Technology, University of York, York, YO10 5DD, United Kingdom.
  • School of Medicine, University of Dundee, Dundee, DD1 9SY, United Kingdom.
  • Institute for Computer Assisted Surgery, University Leipzig, Semmelweisstraße 14, 04103 Leipzig, Germany.

ORCID-Profile: Han Li

Zitieren

APA

Li, H., Barnard, I., Halliwell, T., Zhang, X., Melzer, A., & Huang, Z. (2025). Effects of skull properties on continuous-wave transcranial focused ultrasound transmission. The Journal of the Acoustical Society of America, 157(4), 2336-2349. https://doi.org/10.1121/10.0036344

Vancouver

Li H, Barnard I, Halliwell T, Zhang X, Melzer A, Huang Z. Effects of skull properties on continuous-wave transcranial focused ultrasound transmission. The Journal of the Acoustical Society of America. 2025;157(4):2336-2349. doi:10.1121/10.0036344

BibTeX
@article{Li2025,
  author = {Li, Han and Barnard, Isla and Halliwell, Tyler and Zhang, Xinyu and Melzer, Andreas and Huang, Zhihong},
  title = {Effects of skull properties on continuous-wave transcranial focused ultrasound transmission},
  journal = {The Journal of the Acoustical Society of America},
  year = {2025},
  volume = {157},
  number = {4},
  pages = {2336--2349},
  doi = {10.1121/10.0036344}
}
Kostenlos starten →