Fachartikel in der Zeitschrift Journal of Craniovertebral Junction and Spine, online erschienen am 20.11.2025. Verfasst von Abhishek Soni und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Spine Surgery, Manipal Institute of Robotic Spine Surgery, Manipal Hospital, Bengaluru, Karnataka, India. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Study Design: A single-center prospective study evaluating the accuracy and factors influencing robotic-assisted pedicle screw placement in the thoracolumbar spine.
Objectives
To assess the accuracy of robotic-assisted pedicle screw placement in thoracolumbar spine surgeries and to analyze the factors that enhance or hinder the precision of robotic systems in a single-center setting. Summary of Background Data: Robotic systems are significant advancement in spinal surgery, offering added advantage in pedicle screw placement compared to conventional methods such as freehand, fluoroscopy-guided, and computer-aided navigation (CAN)-guided techniques. Robots combine CAN with a stable mechanical arm, ensuring accurate placement along preplanned trajectories, particularly advantageous in complex anatomies.
Methods
A total of 410 patients who underwent robotic-assisted thoracolumbar spine surgery were included in the study. Pedicle screws were placed with robotic assistance using an optimized workflow. Screw placement accuracy was evaluated using the Gertzbein–Robbins classification, with screws graded A and B considered clinically acceptable.
Results
Of the 2600 screws placed, 99.2% were clinically acceptable (93.4% Grade A and 5.8% Grade B), with only 0.8% exhibiting breaches requiring revision. Lateral breaches were the most common (59.1%). Robotic system usage averaged 20.6 min, with an average time of 3.8 min per screw insertion. Postoperative outcomes included a mean Visual Analog Scale pain score of 7.3 and an average hospital stay of 4.7 days.
Conclusions
Robotic-assisted pedicle screw placement using the Mazor X system demonstrated high accuracy and minimal revision rates. Robotic integration reduces complications and streamlines workflows, improving patient safety and advancing spine surgery standards.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Spine Surgery, Manipal Institute of Robotic Spine Surgery, Manipal Hospital, Bengaluru, Karnataka, India.
- Department of Spine Anesthesia, Manipal Institute of Robotic Spine Surgery, Manipal Hospital, Bengaluru, Karnataka, India.
Zitieren
APA
Soni, A., Srinivasa, V., Joseph, A. X., Thirugnanam, B., & Vidyadhara, A. (2025). Evaluating accuracy in robotic-assisted thoracolumbar pedicle screw placement: Insights from a single-center study of 410 patients. Journal of Craniovertebral Junction and Spine, 16(4), 408-416. https://doi.org/10.4103/jcvjs.jcvjs_134_25
Vancouver
Soni A, Srinivasa V, Joseph AX, Thirugnanam B, Vidyadhara A. Evaluating accuracy in robotic-assisted thoracolumbar pedicle screw placement: Insights from a single-center study of 410 patients. Journal of Craniovertebral Junction and Spine. 2025;16(4):408-416. doi:10.4103/jcvjs.jcvjs_134_25
BibTeX
@article{Soni2025,
author = {Soni, Abhishek and Srinivasa, Vidyadhara and Joseph, Akhil Xavier and Thirugnanam, Balamurugan and Vidyadhara, Alia},
title = {Evaluating accuracy in robotic-assisted thoracolumbar pedicle screw placement: Insights from a single-center study of 410 patients},
journal = {Journal of Craniovertebral Junction and Spine},
year = {2025},
volume = {16},
number = {4},
pages = {408--416},
doi = {10.4103/jcvjs.jcvjs_134_25}
}
EUVIMED