Comparative evaluation of cyclic fatigue resistance of thermomechanically treated NiTi rotary instruments in simulated curved canals with two different radii of curvature: An in vitro Study
Journal of Conservative Dentistry and Endodontics · 2024-04 · 27 (4) · 393-399
DOI: 10.4103/jcde.jcde_32_24 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Conservative Dentistry and Endodontics, online erschienen am 05.04.2024. Verfasst von Tahira Hamid und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Conservative Dentistry and Endodontics, Indira Gandhi Government Dental College, Jammu, Jammu and Kashmir, India. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
This study aimed to compare the effect of heat treatment on the cyclic fatigue resistance of three different nickel–titanium rotary file systems: TruNatomy, HyFlex CM, and NeoEndo flex in simulated curved canals at two different radii of curvatures. Materials and Methods: A total of 60 NiTi rotary files of three types were used, and the samples were divided into three groups TruNatomy, HyFlex CM, and NeoEndo flex. All the instruments were subjected to cyclic fatigue testing using a stainless steel custom-made canal model with a 60° angle of curvature and two radii of curvature 5 mm and 3 mm; the number of cycles to fracture and length of the fractured segment was measured. The separated instruments were subjected to fractographic analysis under scanning electron microscope.
Results
The number of cycles to failure to fracture a file at a 3 mm radius of curvature is significantly less as compared to a 5 mm radius of curvature (HCM > TRN > NE).
Conclusion
Within the limitation of the present study, there was a positive correlation between the radius of curvature and fatigue life of NiTi files.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Conservative Dentistry and Endodontics, Indira Gandhi Government Dental College, Jammu, Jammu and Kashmir, India.
Zitieren
APA
Hamid, T., Malik, A., Kumar, A., & Anjum, S. (2024). Comparative evaluation of cyclic fatigue resistance of thermomechanically treated NiTi rotary instruments in simulated curved canals with two different radii of curvature: An in vitro Study. Journal of Conservative Dentistry and Endodontics, 27(4), 393-399. https://doi.org/10.4103/jcde.jcde_32_24
Vancouver
Hamid T, Malik A, Kumar A, Anjum S. Comparative evaluation of cyclic fatigue resistance of thermomechanically treated NiTi rotary instruments in simulated curved canals with two different radii of curvature: An in vitro Study. Journal of Conservative Dentistry and Endodontics. 2024;27(4):393-399. doi:10.4103/jcde.jcde_32_24
BibTeX
@article{Hamid2024,
author = {Hamid, Tahira and Malik, Azhar and Kumar, Ajay and Anjum, Shamim},
title = {Comparative evaluation of cyclic fatigue resistance of thermomechanically treated NiTi rotary instruments in simulated curved canals with two different radii of curvature: An in vitro Study},
journal = {Journal of Conservative Dentistry and Endodontics},
year = {2024},
volume = {27},
number = {4},
pages = {393--399},
doi = {10.4103/jcde.jcde_32_24}
}