Clinical Evaluation of AI100 with ShonitTM versus Manual Microscopy

C S Jayaprakash, Sandeep Rao, Resmi P Nair, Blessy Mary Thomas, P Chandni Bhandary, Hema N Anadure, C Chaitra

Indian Journal of Pathology and Microbiology · 2026

DOI: 10.4103/ijpm.ijpm_134_25 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Indian Journal of Pathology and Microbiology, 2026. Verfasst von C S Jayaprakash und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology, Father Muller Medical College Hospital, Mangalore, Karnataka, India. Hauptthemen laut MEDLINE: Microscopy.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Peripheral blood smear (PBS) is an essential diagnostic test in current medical practice. AI100 with Shonit TM (AI100) is an automatic cell-locating device intended for white blood cell (WBC) differential count, red blood cell (RBC), and platelet morphology precharacterization. We report a multicenter study evaluating the clinical performance of the AI100 with Shonit™ (combined performance after qualified reviewer verification) in PBS analysis, establishing its equivalence to manual microscopy for diagnostic purposes. Materials and Methods: A retrospective multisite study was conducted at four sites (three in India and one in USA) from August 2021 to January 2023. A total of 882 samples were analyzed by two qualified reviewers in each site independently using manual microscopy and AI100 with Shonit TM . The post-review results were compared between AI100 and manual microscopy.

Results

The overall agreement, sensitivity, and specificity for WBC abnormality are 95.03%, 92.68% and 95.41%, respectively. The overall agreement for RBC anisocytosis and poikilocytosis is 94.7% and 92.1%, respectively. For platelet morphology, the overall agreement, sensitivity, and specificity were 96.8%, 97.9%, and 94.6%, respectively. The repeatability and reproducibility of AI100 with Shonit ™ were also evaluated.

Conclusion

The performance comparison study results demonstrate that the AI100 with Shonit TM is substantially equivalent to manual microscopy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MicroscopyHumansRetrospective StudiesSensitivity and SpecificityReproducibility of ResultsErythrocytesIndiaLeukocyte CountLeukocytesBlood PlateletsBlood Cell Count

Stichwörter der Autor:innen

manual microscopyperipheral blood smear

Autor:innen und Institutionen

  • Department of Pathology, Father Muller Medical College Hospital, Mangalore, Karnataka, India.
  • Department of Hematopathology, Triesta Sciences, A Unit of HCG, Bengaluru, Karnataka, India.
  • Department of Pathology, Believers Church Medical College Hospital, Thiruvalla, Kerala, India.
  • Medical Department, Sigtuple Private Limited, Bengaluru, Karnataka, India.

Zitieren

APA

Jayaprakash, C. S., Rao, S., Nair, R. P., Thomas, B. M., Bhandary, P. C., Anadure, H. N., & Chaitra, C. (2026). Clinical Evaluation of AI100 with ShonitTM versus Manual Microscopy. Indian Journal of Pathology and Microbiology. https://doi.org/10.4103/ijpm.ijpm_134_25

Vancouver

Jayaprakash CS, Rao S, Nair RP, Thomas BM, Bhandary PC, Anadure HN, et al. Clinical Evaluation of AI100 with ShonitTM versus Manual Microscopy. Indian Journal of Pathology and Microbiology. 2026. doi:10.4103/ijpm.ijpm_134_25

BibTeX
@article{Jayaprakash2026,
  author = {Jayaprakash, C S and Rao, Sandeep and Nair, Resmi P and Thomas, Blessy Mary and Bhandary, P Chandni and Anadure, Hema N and Chaitra, C},
  title = {Clinical Evaluation of AI100 with ShonitTM versus Manual Microscopy},
  journal = {Indian Journal of Pathology and Microbiology},
  year = {2026},
  doi = {10.4103/ijpm.ijpm_134_25}
}
Kostenlos starten →