EUVIMEDEUVIMED entdecken
MetaanalyseÜbersichtsarbeitVergleichsstudieOpen Access

The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis

Oscar Albis-Donado, Paulina Ramirez-Neria, Nadia Rios-Acosta, Ingeborg Stalmans

Indian Journal of Ophthalmology · 2024

DOI: 10.4103/ijo.ijo_907_23 ↗

Kurz gefasst

Metaanalyse in der Zeitschrift Indian Journal of Ophthalmology, 2024. Verfasst von Oscar Albis-Donado und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Glaucoma, Instituto Mexicano de Oftalmología IAP, Querétaro City, Querétaro, México. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular, Altitude, Glaucoma. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

A meta-analysis found that including atmospheric pressure as altitude in generalized linear models reveals higher differences between Goldmann tonometry and Pascal dynamic contour tonometry at higher altitudes, with the difference increasing in thinner corneas. To examine the difference in intraocular pressure (IOP) measurements by using Goldman applanation tonometry (GAT) and dynamic contour tonometer (DCT) tonometry in published literature and determine the influence of central corneal thickness (CCT), age, and altitude on that difference.

Methods

Articles that compare GAT and DCT were selected for an extensive literature review, and the location and altitude of the research centers were found online. CCT and age were analyzed as covariates, when available.

Results

A total of 157 studies including 24,211 eyes of 20,214 patients were included in the study. The results showed that the difference between DCT and GAT was higher at higher altitudes above sea level and increased with thinner corneas. However, the results were different in eyes with corneal transplants, where altitude and CCT had less influence, and in those post-refractive surgery where age was found to influence the difference. Theoretical correction formulas using altitude, CCT, and age were derived from this meta-analysis, but their accuracy and usefulness in clinical practice need validation.

Conclusion

The findings suggest that there is a higher risk of underestimating IOP when the Goldmann tonometer is used at a higher altitude, particularly in eyes with glaucoma, thinner corneas, or corneal refractive surgery. Further research is needed to validate the accuracy of the correction formulas derived from this meta-analysis in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intraocular PressureTonometry, OcularAltitudeGlaucomaHumansReproducibility of ResultsCorneaAtmospheric Pressure

Autor:innen und Institutionen

  • Department of Glaucoma, Instituto Mexicano de Oftalmología IAP, Querétaro City, Querétaro, México.
  • Department of Ophthalmology, University Hospitals Leuven (UZ Leuven), Leuven, Belgium.

Zitieren

APA

Albis-Donado, O., Ramirez-Neria, P., Rios-Acosta, N., & Stalmans, I. (2024). The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis. Indian Journal of Ophthalmology. https://doi.org/10.4103/ijo.ijo_907_23

Vancouver

Albis-Donado O, Ramirez-Neria P, Rios-Acosta N, Stalmans I. The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis. Indian Journal of Ophthalmology. 2024. doi:10.4103/ijo.ijo_907_23

BibTeX
@article{AlbisDonado2024,
  author = {Albis-Donado, Oscar and Ramirez-Neria, Paulina and Rios-Acosta, Nadia and Stalmans, Ingeborg},
  title = {The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis},
  journal = {Indian Journal of Ophthalmology},
  year = {2024},
  doi = {10.4103/ijo.ijo_907_23}
}
Kostenlos starten →