Long-term outcome of permanent epicardial pacemaker implantation in neonates: Experience from an Indian center
Annals of Pediatric Cardiology · 2024-03 · 17 (2) · 97-100
DOI: 10.4103/apc.apc_37_24 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of Pediatric Cardiology, online erschienen am 20.07.2024. Verfasst von Deepanjan Bhattacharya und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram, Kerala, India. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 9 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
Permanent pacemaker implantation (PPI) in neonates is challenging with respect to indications, device selection, implantation technique, and long-term outcomes. Complex anatomy, the need for long-term pacing with high rates, and a problematic postoperative period are the major problems.
Methods
We prospectively followed up 22 newborns who underwent PPI below 28 days of life at our institute.
Results
The median age at implantation was 2 days (interquartile range 1–9 days), and 9% were born preterm. The average heart rate before implantation was 46.4 ± 7.2 bpm. Maternal lupus antibodies were positive in 8 (36.4%) neonates, whereas 11 (50.0%) had associated congenital heart disease. Nineteen neonates underwent single chamber (VVI) and three underwent dual chamber (DDD) pacemaker implantation. Over a median follow-up of 46 months (range 2–123 months), the average ventricular pacing percentage was 87.5 ± 24.9%, with a stable pacing threshold. Seven children underwent pulse generator replacement due to battery depletion at a median age of 47 months. Pacing-induced ventricular dysfunction was seen in five children at a median age of 23.6 months, and two underwent upgradation to cardiac resynchronization therapy. Overall mortality was 13.6%, all due to tissue hypoperfusion and lactic acidosis in the postimplantation period.
Conclusions
PPI in neonates has a favorable outcome with excellent lead survival. Overall mortality is 13.6%, which is predominantly in the postimplantation period and related to myocardial dysfunction.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram, Kerala, India.
- Department of Cardiovascular and Thoracic Surgery, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram, Kerala, India.
Zitieren
APA
Bhattacharya, D., Namboodiri, N., Nair, K. K. M., Dharan, B. S., Sasikumar, D., Gopalakrishnan, A., Krishnamoorthy, K. M., Menon, S., Ramanan, S., & Baruah, S. D. (2024). Long-term outcome of permanent epicardial pacemaker implantation in neonates: Experience from an Indian center. Annals of Pediatric Cardiology, 17(2), 97-100. https://doi.org/10.4103/apc.apc_37_24
Vancouver
Bhattacharya D, Namboodiri N, Nair KKM, Dharan BS, Sasikumar D, Gopalakrishnan A, et al. Long-term outcome of permanent epicardial pacemaker implantation in neonates: Experience from an Indian center. Annals of Pediatric Cardiology. 2024;17(2):97-100. doi:10.4103/apc.apc_37_24
BibTeX
@article{Bhattacharya2024,
author = {Bhattacharya, Deepanjan and Namboodiri, Narayanan and Nair, Krishna Kumar Mohanan and Dharan, Baiju S and Sasikumar, Deepa and Gopalakrishnan, Arun and Krishnamoorthy, K M and Menon, Sabarinath and Ramanan, Sowmya and Baruah, Sudip Dutta},
title = {Long-term outcome of permanent epicardial pacemaker implantation in neonates: Experience from an Indian center},
journal = {Annals of Pediatric Cardiology},
year = {2024},
volume = {17},
number = {2},
pages = {97--100},
doi = {10.4103/apc.apc_37_24}
}
EUVIMED