Comparision of Cardiac Output Measured by Transthoracic Echocardiography with Continuous Cardiac Output Measured by Pressure Recording Analytical Method
Annals of Cardiac Anaesthesia · 2023
DOI: 10.4103/aca.aca_41_22 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Annals of Cardiac Anaesthesia, 2023. Verfasst von M Roja Madhuri und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesisiology, SJICR, Bengaluru, Karnataka, India. Hauptthemen laut MEDLINE: Echocardiography, Cardiac Output, Low. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Low cardiac output is a common complication following cardiac surgery and it is associated with higher mortality in the pediatric population. A gold standard method for cardiac output (CO) monitoring in the pediatric population is lacking. The present study was conducted to validate cardiac output and cardiac index measured by transthoracic echocardiography and Pressure recording analytical method, a continuous pulse contour method, MostCare Up in postoperative pediatric cardiac surgical patients. Materials and Methods: This was a prospective observational clinical study conducted at a tertiary care hospital. A total of 23 pediatric patients weighed between 2 and 20 kg who had undergone elective cardiac surgery were included in the study.
Results
Spearman’s correlation coefficient of CO between transthoracic echocardiography (TTE) and Pressure Recording Analytical Method (PRAM) showed of positive correlation (r = 0.69, 95% Confidence interval 0.59-0.77, P < 0.0001) Linear regression equations for CO between TTE and PRAM were y = 0.55 + 0.88x (R 2 = 0.46, P < 0.0001). (y = PRAM, x = TTE), respectively. Bland- Altman plot for CO between TTE and PRAM showed a bias of -0.397 with limits of the agreement being -2.01 to 1.22. Polar plot analysis showed an angular bias of 6.55° with radial limits of the agreement being -21.46 to 34.58 for CO and angular bias of 6.22° with radial limits of the agreement being -22.4 to 34.84 for CI.
Conclusion
PRAM has shown good trending ability for cardiac output. However, values measured by PRAM are not interchangeable with the values measured by transthoracic echocardiography.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anaesthesisiology, SJICR, Bengaluru, Karnataka, India.
- Associate Professor, The Oxford Medical College and Hospital, Bangalore, India.
Zitieren
APA
Madhuri, M. R., Singh, N. G., Nagaraja, P. S., Pranathi, E., Prabhu Shankar, C. G., Satish, N., Prabhakar, V., & Gupta, A. (2023). Comparision of Cardiac Output Measured by Transthoracic Echocardiography with Continuous Cardiac Output Measured by Pressure Recording Analytical Method. Annals of Cardiac Anaesthesia. https://doi.org/10.4103/aca.aca_41_22
Vancouver
Madhuri MR, Singh NG, Nagaraja PS, Pranathi E, Prabhu Shankar CG, Satish N, et al. Comparision of Cardiac Output Measured by Transthoracic Echocardiography with Continuous Cardiac Output Measured by Pressure Recording Analytical Method. Annals of Cardiac Anaesthesia. 2023. doi:10.4103/aca.aca_41_22
BibTeX
@article{Madhuri2023,
author = {Madhuri, M Roja and Singh, Naveen G and Nagaraja, P S and Pranathi, E and Prabhu Shankar, C G and Satish, N and Prabhakar, V and Gupta, Apoorva},
title = {Comparision of Cardiac Output Measured by Transthoracic Echocardiography with Continuous Cardiac Output Measured by Pressure Recording Analytical Method},
journal = {Annals of Cardiac Anaesthesia},
year = {2023},
doi = {10.4103/aca.aca_41_22}
}
EUVIMED