Accuracy and precision of ultrasonic cardiac output monitor versus echocardiography for cardiac output measurement in neonates: a systematic review and meta-analysis

Na Wang, Bin Tang, Hongjuan Zhang, Bo Wang, Jia Zhang

Medical Ultrasonography · 2025

DOI: 10.11152/mu-4538 ↗

Kurz gefasst

Systematische Übersicht in der Zeitschrift Medical Ultrasonography, 2025. Verfasst von Na Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neonatology, The Affiliated Suqian First People's Hospital of Nanjing Medical University, Suqian, Jiangsu. Hauptthemen laut MEDLINE: Cardiac Output, Echocardiography.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

To perform a systematic review and meta-analysis of the consistency between ultrasonic cardiac output monitor (USCOM) and echocardiography in monitoring neonatal cardiac output.Material and methods: PubMed, Embase, Web of Science, Cochrane, Wanfang, Chinese National Knowledge Infrastructure Database, and Chinese Science and Technique Journal Database were searched from the inception of the databases to November 2024. Two reviewers independently performed study selection, data extraction, and risk of bias assessment. Method-comparison studies evaluating USCOM versus echocardiography in neonates were included. Pooled mean bias (accuracy), standard deviation, and mean percentage error (precision) of the included studies were calculated using a fixed-effects model.

Results

Six of the 774 identified studies were included in the qualitative analysis, and four studies were included in the meta-analysis. The overall pooled bias and percentage error for left ventricular output measurements were -7.7±13.5 mL/kg/min (95% limit of agreement: -34.2 to 18.8 mL/kg/min) and 31.2%, respectively, while the overall pooled bias and percentage error for right ventricular output measurements were 40.5±21.3 mL/kg/min (95% limit of agreement: -1.2 to 82.2 mL/kg/min) and 35.1%, respectively.

Conclusion

USCOM may overestimate right ventricular output trend, while left ventricular output measurement agreement is slightly better.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac OutputEchocardiographyHumansInfant, NewbornReproducibility of ResultsMonitoring, Physiologic

Autor:innen und Institutionen

  • Department of Neonatology, The Affiliated Suqian First People's Hospital of Nanjing Medical University, Suqian, Jiangsu.
  • Department of Neonatology, The Affiliated Suqian First People's Hospital of Nanjing Medical University, Suqian, Jiangsu. ZJsuqian8383@126.com.

Zitieren

APA

Wang, N., Tang, B., Zhang, H., Wang, B., & Zhang, J. (2025). Accuracy and precision of ultrasonic cardiac output monitor versus echocardiography for cardiac output measurement in neonates: a systematic review and meta-analysis. Medical Ultrasonography. https://doi.org/10.11152/mu-4538

Vancouver

Wang N, Tang B, Zhang H, Wang B, Zhang J. Accuracy and precision of ultrasonic cardiac output monitor versus echocardiography for cardiac output measurement in neonates: a systematic review and meta-analysis. Medical Ultrasonography. 2025. doi:10.11152/mu-4538

BibTeX
@article{Wang2025,
  author = {Wang, Na and Tang, Bin and Zhang, Hongjuan and Wang, Bo and Zhang, Jia},
  title = {Accuracy and precision of ultrasonic cardiac output monitor versus echocardiography for cardiac output measurement in neonates: a systematic review and meta-analysis},
  journal = {Medical Ultrasonography},
  year = {2025},
  doi = {10.11152/mu-4538}
}
Kostenlos starten →