Optimal Plate Position for Biomechanical Stability in Medial Opening-Wedge High Tibial Osteotomy: A Finite Element Analysis

Hyun-Soo Moon iD, Jin-Ho Youn iD, Sung-Jae Lee iD, Dae-Kyung Kwak iD, Kwang-Min Park iD, Shi-Hyun Kim iD, Je-Hyun Yoo iD

Clinics in Orthopedic Surgery · 2025 · 17 (4) · 622

DOI: 10.4055/cios24431 ↗

At a glance

Journal article in Clinics in Orthopedic Surgery, 2025. Written by Hyun-Soo Moon and 6 co-authors. Institution: Department of Orthopedic Surgery, Gangnam Severance Hospital Hospital, Yonsei University College of Medicine, Seoul, Korea. Major MEDLINE topics: Tibia, Osteotomy, Bone Plates. Cited 3 times so far (Crossref); 28 references. The full text is freely available (open access, CC BY-NC 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Research on the ideal fixation position for plates in medial opening-wedge high tibial osteotomy (MOWHTO) directly applicable in clinical settings is scarce. Therefore, this study aimed to evaluate the biomechanical effects of different plate positions in MOWHTO through finite element analysis (FEA) to explore a potentially optimal plate position.

Methods

Utilizing the computed tomography images of a 67-year-old man, a 3-dimensional model of the knee, along with an implant (TomoFix standard plate and screws), was created to simulate a virtual MOWHTO with a 10° medial opening gap. Biomechanical stability analysis of the bone-implant construct was conducted through FEA under physiologic loading simulating a 1-legged stance, with varying plate positions. Configurations for plate fixation, determined by anterior-posterior depth and height, resulted in a total of 9 fixation positions (anterior, center, and posterior in terms of depth; proximal, middle, and distal in terms of height). Criteria for assessment included inter-fragmentary micromotion at the medial opening gap, mean stress on the lateral hinge, the entire tibial bone, and the implant, stress shielding effect, and peak von Mises stress (PVMS).

Results

The inter-fragmentary micromotion at the medial opening gap exhibited a tendency to decrease as the fixation position of the plate moved posteriorly and proximally, observed in both axial and shear micromotion. The mean stress on the lateral hinge of the tibia progressively decreased with more posterior or proximal plate placement, reaching its minimum in the most posterior and proximal position. In terms of the mean stress imposed on both the entire bone and implant, it decreased when the plate was positioned posteriorly and proximally, and this position was deemed favorable from the perspective of the stress shielding effect. PVMS predominantly occurred at hole 1 of the plate and its corresponding screw, and it was lower than the yield strength of the titanium alloy regardless of the plate's position.

Conclusions

Placing the plate more posteriorly and proximally in MOWHTO could minimize inter-fragmentary micromotion, reduce stress on the lateral hinge and bone-implant construct, and enhance stress shielding, all without increasing the risk of implant breakage, suggesting it as a potentially optimal plate position.

Subjects

MeSH terms (MEDLINE)

TibiaOsteotomyBone PlatesHumansFinite Element AnalysisAgedBiomechanical PhenomenaMaleTomography, X-Ray ComputedImaging, Three-Dimensional

Author keywords

High tibial osteotomyMicromotionPlate positionStabilityStress

Authors and institutions

  • Department of Orthopedic Surgery, Gangnam Severance Hospital Hospital, Yonsei University College of Medicine, Seoul, Korea.
  • Department of Biomedical Engineering, Inje University, Gimhae, Korea.
  • Department of Orthopedic Surgery, Hallym University Sacred Heart Hospital, Hallym University College of Medicine, Anyang, Korea.
  • Medical Device Development Center, Osong Medical Innovation Foundation, Cheongju, Korea.

ORCID profiles: Hyun-Soo Moon, Jin-Ho Youn, Sung-Jae Lee, Dae-Kyung Kwak, Kwang-Min Park, Shi-Hyun Kim, Je-Hyun Yoo

Cite

APA

Moon, H. S., Youn, J. H., Lee, S. J., Kwak, D. K., Park, K. M., Kim, S. H., & Yoo, J. H. (2025). Optimal Plate Position for Biomechanical Stability in Medial Opening-Wedge High Tibial Osteotomy: A Finite Element Analysis. Clinics in Orthopedic Surgery, 17(4), 622. https://doi.org/10.4055/cios24431

Vancouver

Moon HS, Youn JH, Lee SJ, Kwak DK, Park KM, Kim SH, et al. Optimal Plate Position for Biomechanical Stability in Medial Opening-Wedge High Tibial Osteotomy: A Finite Element Analysis. Clinics in Orthopedic Surgery. 2025;17(4):622. doi:10.4055/cios24431

BibTeX
@article{Moon2025,
  author = {Moon, Hyun-Soo and Youn, Jin-Ho and Lee, Sung-Jae and Kwak, Dae-Kyung and Park, Kwang-Min and Kim, Shi-Hyun and Yoo, Je-Hyun},
  title = {Optimal Plate Position for Biomechanical Stability in Medial Opening-Wedge High Tibial Osteotomy: A Finite Element Analysis},
  journal = {Clinics in Orthopedic Surgery},
  year = {2025},
  volume = {17},
  number = {4},
  pages = {622},
  doi = {10.4055/cios24431}
}
Start for free →