High-Performance Graphene-Modified Sensing Chip for SARS-CoV-2 Detection
Journal of Visualized Experiments · 2023
DOI: 10.3791/64730 ↗At a glance
Video/Audio in Journal of Visualized Experiments, 2023. Written by Parshant Kumar Sharma and 4 co-authors. Institution: RFIC Bio Centre, Kwangwoon University; Department of Electronics Engineering, Kwangwoon University; NDAC Centre, Kwangwoon University; parshantvats11…. Major MEDLINE topics: COVID-19, Graphite, Viruses, Biosensing Techniques. Cited 6 times so far (Crossref); 50 references. Funded by: NIBIB NIH HHS.
Abstract (original language)
This sensing prototype model involves the development of a reusable, twofold graphene oxide (GrO)-glazed double inter-digitated capacitive (DIDC) detecting chip for detecting severe acute respiratory syndrome coronavirus 2 virus (SARS-CoV-2) specifically and rapidly. The fabricated DIDC comprises a Ti/Pt-containing glass substrate glazed with graphene oxide (GrO), which is further chemically modified with EDC-NHS to immobilize antibodies (Abs) hostile to SARS-CoV-2 based on the spike (S1) protein of the virus. The results of insightful investigations showed that GrO gave an ideal engineered surface for Ab immobilization and enhanced the capacitance to allow higher sensitivity and low sensing limits. These tunable elements helped accomplish a wide sensing range (1.0 mg/mL to 1.0 fg/mL), a minimum sensing limit of 1 fg/mL, high responsiveness and good linearity of 18.56 nF/g, and a fast reaction time of 3 s. Besides, in terms of developing financially viable point-of-care (POC) testing frameworks, the reusability of the GrO-DIDC biochip in this study is good. Significantly, the biochip is specific against blood-borne antigens and is stable for up to 10 days at 5 °C. Due to its compactness, this scaled-down biosensor has the potential for POC diagnostics of COVID-19 infection. This system can also detect other severe viral diseases, although an approval step utilizing other virus examples is under development.
Subjects
MeSH terms (MEDLINE)
Substances
Authors and institutions
- RFIC Bio Centre, Kwangwoon University; Department of Electronics Engineering, Kwangwoon University; NDAC Centre, Kwangwoon University; parshantvats111@gmail.com.
- Stanford Cardiovascular Institute, Stanford University School of Medicine; Department of Medicine, Stanford University School of Medicine; ebi.mostafavi@gmail.com.
- RFIC Bio Centre, Kwangwoon University; Department of Electronics Engineering, Kwangwoon University; NDAC Centre, Kwangwoon University; nykim@kw.ac.kr.
- Department of Biomedical Engineering, Hebei University of Technology; websterthomas02@gmail.com.
- NanoBioTech Laboratory, Department of Environmental Engineering, Florida Polytechnic University; School of Engineering, University of Petroleum and Energy Studies (UPES); akaushik@floridapoly.edu.
Funding
- NIBIB NIH HHS
Cite
APA
Sharma, P. K., Mostafavi, E., Kim, N. Y., Webster, T. J., & Kaushik, A. (2023). High-Performance Graphene-Modified Sensing Chip for SARS-CoV-2 Detection. Journal of Visualized Experiments. https://doi.org/10.3791/64730
Vancouver
Sharma PK, Mostafavi E, Kim NY, Webster TJ, Kaushik A. High-Performance Graphene-Modified Sensing Chip for SARS-CoV-2 Detection. Journal of Visualized Experiments. 2023. doi:10.3791/64730
BibTeX
@article{Sharma2023,
author = {Sharma, Parshant Kumar and Mostafavi, Ebrahim and Kim, Nam-Young and Webster, Thomas J and Kaushik, Ajeet},
title = {High-Performance Graphene-Modified Sensing Chip for SARS-CoV-2 Detection},
journal = {Journal of Visualized Experiments},
year = {2023},
doi = {10.3791/64730}
}
EUVIMED