EUVIMEDEUVIMED entdecken
Crossref

Finite element analysis of strain effects at bone – orthodontic mini-implant interface

Mihai A. Niculescu, Cătălin Popescu, Dan Păunescu

Romanian Journal of Stomatology · 2015

DOI: 10.37897/rjs.2015.1.2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Romanian Journal of Stomatology, 2015. Verfasst von Mihai A. Niculescu und 2 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The study analyses the peri-implant bone strain resulted during implant anchorage for molar intrusion. Material and method. A model and a mini-implant in finite element were created using tridimensional scanning and volumetric computer tomograf, and used to simulate molar intrusion using mini-implant anchorage. The force of 224 g was applied on the first molar and 122 g on each mini-implant. ANSYSv11 program was used to measure the stress-strain relation in the bone model using finite element analysis. Results and discussion. The peri-implant bone maximum stress value of 9.8 MPa correspond to 890 microstrain and molar peri-radicular stress value of 2.5 MPa correspond to 36 microstrain.

Conclusions

Finite element analysis is a complex mathematical calculation allowing to receive and process data in the model, data that can be transferred in clinical practice. The data resulted in this study, correlated with the estimated effects of bone remodeling effects when subjected to mechanical factors, shows that the mini-implant mechanical induced values of peri-implant and periradicular bone strain are located in-between disuse interval and adapted interval. In order to increase the relevance of the results, an improved computer tomographic investigation and calibrated model related to the patient characteristics, are needed.

Zitieren

APA

Niculescu, M. A., Popescu, C., & Păunescu, D. (2015). Finite element analysis of strain effects at bone – orthodontic mini-implant interface. Romanian Journal of Stomatology. https://doi.org/10.37897/rjs.2015.1.2

Vancouver

Niculescu MA, Popescu C, Păunescu D. Finite element analysis of strain effects at bone – orthodontic mini-implant interface. Romanian Journal of Stomatology. 2015. doi:10.37897/rjs.2015.1.2

BibTeX
@article{Niculescu2015,
  author = {Niculescu, Mihai A. and Popescu, Cătălin and Păunescu, Dan},
  title = {Finite element analysis of strain effects at bone – orthodontic mini-implant interface},
  journal = {Romanian Journal of Stomatology},
  year = {2015},
  doi = {10.37897/rjs.2015.1.2}
}
Kostenlos starten →