Quantifying Patient Positioning Errors and Radiation Dose Variation in Cardiac Computed Tomography Angiography

Yasser H Hadi, Hassan B Hawsawi, Niamh Moore, Helle Precht, Andrew England, Mark McEntee

Journal of the Saudi Heart Association · 08.05.2026 · 38 (2)

DOI: 10.37616/2212-5043.1499 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of the Saudi Heart Association, online erschienen am 08.05.2026. Verfasst von Yasser H Hadi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Discipline of Medical Imaging and Radiation Therapy, School of Medicine and Health, University College Cork, Cork, Ireland. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Accurate patient centring at the computed tomography (CT) scanner isocentre is essential for optimising radiation dose. This study aimed to quantify positioning errors, assess their relationship with radiation dose, and estimate the potential dose reduction achievable with optimal centring.

Methods

This retrospective observational study included adult patients (≥18 years) who underwent electrocardiography (ECG)-gated cardiac CT angiography between January 2023 and July 2025 at a tertiary hospital. Scans were performed using Siemens SOMATOM Force dual-source CT systems. Vertical and lateral deviations from the scanner isocentre were calculated from scout images using integrated analytics within the institutional dose monitoring platform (Qaelum NV, Leuven, Belgium). Radiation dose indices, including the volume-weighted CT dose index (CTDIvol) and dose length product (DLP), were extracted from DICOM dose reports. Statistical analyses comprised descriptive statistics, correlation analysis, multivariable regression, and simulation of potential dose reductions with ideal centring.

Results

Data from 399 patients were evaluated. Significant off-centring (>10 mm) occurred in 83% of scans in the anteroposterior (AP) direction and 50% laterally. Severe AP off-centring (>30 mm) was present in 39% of cases. Off-centring showed no independent association with CTDIvol or DLP after adjustment for body mass index (BMI), age, and sex. Male patients demonstrated significantly greater off-centring in both the AP (p=0.048) and lateral (p=0.037) directions, and BMI category was significantly associated with off-centring magnitude in the AP (p=0.031) and lateral (p=0.022) directions. Literature-based scenario modelling suggested that ideal centring could hypothetically reduce mean CTDIvol by approximately 12%, with a similar reduction in DLP.

Conclusion

Patient off-centring was highly prevalent in cardiac CT examinations. No independent association between off-centring and radiation dose was observed when adjusting for BMI, age, and sex. Literature-based scenario modelling suggests that meaningful radiation dose reductions may be achievable with optimum centring, although this requires prospective confirmation.

Schlagwörter

Stichwörter der Autor:innen

CTCardiac CTOff-centringPatient positioningRadiation dose

Autor:innen und Institutionen

  • Discipline of Medical Imaging and Radiation Therapy, School of Medicine and Health, University College Cork, Cork, Ireland.
  • Department of Medical Imaging and Intervention, King Abdullah Medical City (KAMC), Makkah, Saudi Arabia.
  • Department of Medical Physics, King Abdullah Medical City (KAMC), Makkah, Saudi Arabia.

Zitieren

APA

Hadi, Y. H., Hawsawi, H. B., Moore, N., Precht, H., England, A., & McEntee, M. (2026). Quantifying Patient Positioning Errors and Radiation Dose Variation in Cardiac Computed Tomography Angiography. Journal of the Saudi Heart Association, 38(2). https://doi.org/10.37616/2212-5043.1499

Vancouver

Hadi YH, Hawsawi HB, Moore N, Precht H, England A, McEntee M. Quantifying Patient Positioning Errors and Radiation Dose Variation in Cardiac Computed Tomography Angiography. Journal of the Saudi Heart Association. 2026;38(2). doi:10.37616/2212-5043.1499

BibTeX
@article{Hadi2026,
  author = {Hadi, Yasser H and Hawsawi, Hassan B and Moore, Niamh and Precht, Helle and England, Andrew and McEntee, Mark},
  title = {Quantifying Patient Positioning Errors and Radiation Dose Variation in Cardiac Computed Tomography Angiography},
  journal = {Journal of the Saudi Heart Association},
  year = {2026},
  volume = {38},
  number = {2},
  doi = {10.37616/2212-5043.1499}
}
Kostenlos starten →