Ultra-low-dose coronary computed tomography angiography using photon-counting detector computed tomography

Suguru Araki iD, Satoshi Nakamura iD, Masafumi Takafuji, Yasutaka Ichikawa, Hajime Sakuma, Kakuya Kitagawa

European Heart Journal - Imaging Methods and Practice · 25.07.2024 · 2 (3)

DOI: 10.1093/ehjimp/qyae125 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Heart Journal - Imaging Methods and Practice, online erschienen am 27.11.2024. Verfasst von Suguru Araki und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mie University Hospital, Tsu, Japan. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Photon-counting detector computed tomography (PCD-CT), which allows the exclusion of electronic noise, shows promise for significant dose reduction in coronary CT angiography (CCTA). This study aimed to assess the radiation dose and image quality of CCTA using PCD-CT, combined with high-pitch helical scanning and an ultra-low tube potential of 70 kVp, and investigate the effect of a sharp kernel on image quality and stenosis assessment in such an ultra-low-dose CCTA setting. Methods and results Forty patients (65% male) with stable heart rates and no prior coronary interventions were included. Data on CT dose index volume (CTDIvol) and dose-length product (DLP) were collected, with effective radiation dose estimated using a conversion factor of 0.014. Images were reconstructed using kernels of Bv64 and Bv40 for image quality and stenosis assessment. The mean CTDIvol, DLP, and effective dose of CCTA were 1.72 ± 0.38 mGy, 29.1 ± 6.8 mGy·cm, and 0.41 ± 0.09 mSv, respectively. Image quality was similar (P = 0.75) between the two kernels, with over 95% of segments achieving a rating of good image quality for both kernels. The per-segment stenosis score distribution between Bv40 and Bv64 reconstruction images showed significant differences for both non-calcified and calcified plaques (P < 0.001 for both).

Conclusion

PCD-CT technology with high-pitch helical scanning and the tube potential of 70 kVp can provide CCTA with ultra-low radiation exposure (DLP, 29 mGy·cm). The noise reduction capability of PCD-CT allows the use of a sharp kernel even in this low-dose CCTA setting without compromising image quality, potentially improving the evaluation of coronary artery stenosis.

Schlagwörter

Stichwörter der Autor:innen

computed tomography angiographycoronary arteriesdual-source CTlow radiation exposurephoton-counting CT

Autor:innen und Institutionen

  • Department of Radiology, Mie University Hospital, Tsu, Japan.
  • Department of Advanced Diagnostic Imaging, Mie University Graduate School of Medicine, 2-174 Edobashi, Tsu, Mie 514-8507, Japan.
  • Regional Co-creation Deployment Center, Mie Regional Plan Co-creation Organization, Mie University, Tsu, Japan.

ORCID-Profile: Suguru Araki, Satoshi Nakamura

Zitieren

APA

Araki, S., Nakamura, S., Takafuji, M., Ichikawa, Y., Sakuma, H., & Kitagawa, K. (2024). Ultra-low-dose coronary computed tomography angiography using photon-counting detector computed tomography. European Heart Journal - Imaging Methods and Practice, 2(3). https://doi.org/10.1093/ehjimp/qyae125

Vancouver

Araki S, Nakamura S, Takafuji M, Ichikawa Y, Sakuma H, Kitagawa K. Ultra-low-dose coronary computed tomography angiography using photon-counting detector computed tomography. European Heart Journal - Imaging Methods and Practice. 2024;2(3). doi:10.1093/ehjimp/qyae125

BibTeX
@article{Araki2024,
  author = {Araki, Suguru and Nakamura, Satoshi and Takafuji, Masafumi and Ichikawa, Yasutaka and Sakuma, Hajime and Kitagawa, Kakuya},
  title = {Ultra-low-dose coronary computed tomography angiography using photon-counting detector computed tomography},
  journal = {European Heart Journal - Imaging Methods and Practice},
  year = {2024},
  volume = {2},
  number = {3},
  doi = {10.1093/ehjimp/qyae125}
}
Kostenlos starten →