Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field
Japanese Journal of Applied Physics · 06.11.2026
DOI: 10.35848/1347-4065/aebf3d ↗Kurz gefasst
Fachartikel in der Zeitschrift Japanese Journal of Applied Physics, online erschienen am 06.11.2026. Verfasst von Liang Zhao und 8 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science.
Abstract (Originalsprache)
Out-of-field neutron dose evaluation is important for radiation protection during boron neutron capture therapy (BNCT). However, no commercial detector is available for measuring the neutron ambient dose equivalent H* outside the irradiation field. Establishing a reliable H* evaluation method is essential for the international standardization of BNCT. In this study, a prototype neutron detector was evaluated using a cyclotron-based epithermal neutron source (C-BENS). The ratio of the measured to calculated H*, based on ICRP Publication 116, ranged from 0.21–0.56 at lateral distances of 10–150 cm from the beam aperture center. This discrepancy is likely due to the difficulty in accurately reproducing the experimental geometry in the Monte Carlo simulation used to calculate the reference H*. These results provide a basis for improving H* evaluation methods and developing standardized radiation protection procedures for BNCT.
Förderung
- Japan Society for the Promotion of Science
Zitieren
APA
Zhao, L., Tanaka, H., Hu, N., Takata, T., Kawabata, S., Kurosawa, S., Yamaji, A., Masuda, A., & MATSUMOTO, T. (2026). Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field. Japanese Journal of Applied Physics. https://doi.org/10.35848/1347-4065/aebf3d
Vancouver
Zhao L, Tanaka H, Hu N, Takata T, Kawabata S, Kurosawa S, et al. Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field. Japanese Journal of Applied Physics. 2026. doi:10.35848/1347-4065/aebf3d
BibTeX
@article{Zhao2026,
author = {Zhao, Liang and Tanaka, Hiroki and Hu, Naonori and Takata, Takushi and Kawabata, Shinji and Kurosawa, Shunsuke and Yamaji, Akihiro and Masuda, Akihiko and MATSUMOTO, Tetsuro},
title = {Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT field},
journal = {Japanese Journal of Applied Physics},
year = {2026},
doi = {10.35848/1347-4065/aebf3d}
}