Vergleichsstudie in der Zeitschrift Tomography, 2026. Verfasst von K Shahgeldi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Therapy, Uppsala University Hospital, 751 85 Uppsala, Sweden. Hauptthemen laut MEDLINE: Radiation Dosage, Pelvimetry, Tomography, Spiral Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The present study aimed to assess the radiation dose associated with low-dose (LD) CT pelvimetry compared with conventional radiography and to evaluate the adequacy of the resulting image quality.
Methods
The absorbed dose was measured using thermoluminescent dosimeters positioned in an anthropomorphic female phantom, including uterine locations, to estimate the fetal dose. Conventional radiographic pelvimetry and LD-CT pelvimetry were performed using clinically implemented protocols. Effective dose was calculated using Monte Carlo–based modeling applying acquisition parameters and retrospective patient dose registry data. Image quality of LD-CT pelvimetry was independently evaluated in 14 consecutive clinical cases using a four-point ordinal scale.
Results
LD-CT pelvimetry reduced the mean absorbed pelvic dose by approximately 50% compared with conventional pelvimetry (0.18 vs. 0.39 mGy) and decreased estimated fetal dose by 40% (0.21 vs. 0.37 mGy). These estimates were based on standardized single acquisitions and did not incorporate additional radiation from retakes commonly observed in conventional practice. CT demonstrated substantially more homogeneous dose distribution, whereas conventional pelvimetry exhibited marked heterogeneity with peak values up to 2.3 mGy. The maternal effective dose was lower for LD-CT (0.16 mSv) than for conventional pelvimetry (0.36 mSv); inclusion of retakes increased the conventional effective dose to 0.71 mSv. All CT examinations were diagnostically adequate, and no recalls were required.
Conclusions
Optimized low-dose CT pelvimetry significantly reduces radiation dose compared with conventional radiographic pelvimetry while maintaining reliable diagnostic image quality. These results support the clinical adoption of CT-based pelvimetry as a dose-efficient and reproducible alternative to conventional techniques.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Therapy, Uppsala University Hospital, 751 85 Uppsala, Sweden.
- Department of Imaging and Functional Medicine, Division Diagnostics, Gävle Hospital, Region Gävleborg, 801 88 Gävle, Sweden.
- Centre for Research and Development, Uppsala University, Region Gävleborg, 801 88 Gävle, Sweden.
ORCID-Profile: T M Svahn
Zitieren
APA
Shahgeldi, K., Parenmark, M., Claesson, L., & Svahn, T. M. (2026). Evaluation of Radiation Dose and Image Quality in the Transition from Conventional Pelvimetry to Low-Dose Helical CT Pelvimetry. Tomography. https://doi.org/10.3390/tomography12030035
Vancouver
Shahgeldi K, Parenmark M, Claesson L, Svahn TM. Evaluation of Radiation Dose and Image Quality in the Transition from Conventional Pelvimetry to Low-Dose Helical CT Pelvimetry. Tomography. 2026. doi:10.3390/tomography12030035
BibTeX
@article{Shahgeldi2026,
author = {Shahgeldi, K and Parenmark, M and Claesson, L and Svahn, T M},
title = {Evaluation of Radiation Dose and Image Quality in the Transition from Conventional Pelvimetry to Low-Dose Helical CT Pelvimetry},
journal = {Tomography},
year = {2026},
doi = {10.3390/tomography12030035}
}
EUVIMED