Fachartikel in der Zeitschrift Tomography, 2025. Verfasst von Zuhal Y Hamd und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riya…. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed, Radiography, Abdominal. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Princess Nourah bint Abdulrahman University.
Abstract (Originalsprache)
Background
As a diagnostic radiology procedure, computed tomography (CT) contributes to patient radiation exposure; hence, it deserves special consideration. The use of diagnostic reference levels (DRLs) is an efficient way to optimize patient radiation dosage. The computed tomography dose index volume (CTDIv) and the dose-length product (DLP) help to measure DRLs.
Methods
A retrospective analysis was conducted on 106 patients (43.9% male, 56.1% female; mean age of 48.18 years) who underwent computed tomography chest, abdomen, and pelvis (CT CAP) scans using a Toshiba Aquilion Prime 160-slice CT scanner. Data included patient demographics, CT parameters (mA, tube rotation time, pitch, slice thickness, and slice count), and dose indices: dose length product (DLP), computed tomography dose index volume (CTDIvol), and effective dose. Cancer risks were calculated based on effective dose, patient demographics, and scan parameters.
Results
This study demonstrated that the mean values for DLP, CTDIvol, and effective dose were 1719.64 ± 488.45 mGy·cm, 25.97 ± 6.96 mGy, and 27.5 ± 7.82 mSv, respectively. Cancer risk estimates ranged from 0.048% to 1.58%, with higher risks observed for females, younger patients. Significant correlations were found between dose indices and technical parameters, including pitch, kVp, tube rotation time, and slice thickness (p < 0.005).
Conclusions
The mean values for DLP, CTDIvol, and effective dose for abdominopelvic scans were higher than those found in previous studies, with significant correlation of weight on these values. Optimizing CT protocols and establishing DRLs tailored to clinical indications are critical for minimizing radiation exposure and enhancing patient safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, P.O. Box 84428, Riyadh 11671, Saudi Arabia.
- Department of Applied Radiologic Technology, College of Applied Medical Sciences, University of Jeddah, Jeddah 23218, Saudi Arabia.
- Department of Radiotherapy, College of Medical Radiologic Sciences, Sudan University of Science and Technology, Khartoum P.O. Box 11111, Sudan.
- Department of Diagnostic Radiology, College of Applied Medical Sciences, Taibah University, Al-Madinah Al-Munawwarah 42353, Saudi Arabia.
ORCID-Profile: Zuhal Y Hamd, Huda I Almohammed, Abdoelrahman Hassan A B, Awadia Gareeballah
Förderung
- Princess Nourah bint Abdulrahman University
Zitieren
APA
Hamd, Z. Y., Almohammed, H. I., Mansour, E., Hassan A B, A., & Gareeballah, A. (2025). Optimizing Radiation Risk Assessment in CT Imaging: Establishing Institutional Diagnostic Reference Levels and Personalized Dose Strategies for Chest, Abdomen, and Pelvis Scans. Tomography. https://doi.org/10.3390/tomography11060065
Vancouver
Hamd ZY, Almohammed HI, Mansour E, Hassan A B A, Gareeballah A. Optimizing Radiation Risk Assessment in CT Imaging: Establishing Institutional Diagnostic Reference Levels and Personalized Dose Strategies for Chest, Abdomen, and Pelvis Scans. Tomography. 2025. doi:10.3390/tomography11060065
BibTeX
@article{Hamd2025,
author = {Hamd, Zuhal Y and Almohammed, Huda I and Mansour, Elbagir and Hassan A B, Abdoelrahman and Gareeballah, Awadia},
title = {Optimizing Radiation Risk Assessment in CT Imaging: Establishing Institutional Diagnostic Reference Levels and Personalized Dose Strategies for Chest, Abdomen, and Pelvis Scans},
journal = {Tomography},
year = {2025},
doi = {10.3390/tomography11060065}
}
EUVIMED