Convolutional Neural Network-Based Alpha/Beta Pulse Shape Discrimination for Low-Energy Tritium Monitoring in Liquid Scintillation Counting
Technologies · 10.06.2026 · 14 (6) · 349
DOI: 10.3390/technologies14060349 ↗Kurz gefasst
Fachartikel in der Zeitschrift Technologies, online erschienen am 10.06.2026. Verfasst von Jie Ren und 4 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Chinese Postdoctoral Science Foundation, Fundamental Research Funds for the Central Universities.
Abstract (Originalsprache)
Alpha/beta (α/β) pulse shape discrimination (PSD) in liquid scintillation counting (LSC) is fundamentally limited by the charge comparison method (CCM) at low energies, where the entire tritium (3H) beta spectrum resides (0–18.6 keVee). The CCM figure-of-merit drops below 0.6 in this region, rendering it inadequate for simultaneous tritium and natural uranium alpha monitoring in nuclear power plant (NPP) liquid effluents. We present a one-dimensional convolutional neural network (1D-CNN) trained on an 80,000-waveform physics-based simulation dataset using established scintillation parameters for Ultima Gold AB. The proposed network achieves 97.4% overall classification accuracy and an area under the receiver operating characteristic curve (AUC) of 0.9981 on the held-out test set, representing improvements of 13.8 percentage points and 0.046 AUC over CCM. In the critical 0–18.6 keVee region, CNN accuracy exceeds 95% compared to below 60% for CCM—a greater than 35 percentage point improvement. Pulse amplitude discrimination (PAD), evaluated as a preliminary screening method, exhibits a 6.3% alpha spillover rate into the beta window, exceeding the regulatory limit of 3%. Gradient-weighted class activation maps (Grad-CAM) confirm that the network exploits physically meaningful pulse features rather than simulation artefacts. A comprehensive background suppression strategy combining dual-SiPM coincidence (24× reduction), anti-coincidence guard detector (5.8× reduction), composite passive shielding (10× reduction), and CNN-assisted discrimination reduces the system equivalent background to 1.83 ± 0.12 cpm, yielding a tritium minimum detectable activity (MDA) of 0.21 Bq/mL (10 mL sample, 30 min count), which satisfies the GB 14587 reference limit of 0.5 Bq/mL. After 8-bit post-training quantisation, the model achieves sub-microsecond inference latency on an embedded Xilinx Artix-7 Field-programmable gate array(FPGA), enabling real-time deployment in portable online monitoring systems.
Förderung
- National Natural Science Foundation of China
- Chinese Postdoctoral Science Foundation
- Fundamental Research Funds for the Central Universities
- Taizhou Science and Technology Support Program Project
- Lianyungang Basic Research Program
Zitieren
APA
Ren, J., Wang, P., Gu, A., Gong, C., & Yang, Y. (2026). Convolutional Neural Network-Based Alpha/Beta Pulse Shape Discrimination for Low-Energy Tritium Monitoring in Liquid Scintillation Counting. Technologies, 14(6), 349. https://doi.org/10.3390/technologies14060349
Vancouver
Ren J, Wang P, Gu A, Gong C, Yang Y. Convolutional Neural Network-Based Alpha/Beta Pulse Shape Discrimination for Low-Energy Tritium Monitoring in Liquid Scintillation Counting. Technologies. 2026;14(6):349. doi:10.3390/technologies14060349
BibTeX
@article{Ren2026,
author = {Ren, Jie and Wang, Peng and Gu, Ao-Tian and Gong, Chunhui and Yang, Yi},
title = {Convolutional Neural Network-Based Alpha/Beta Pulse Shape Discrimination for Low-Energy Tritium Monitoring in Liquid Scintillation Counting},
journal = {Technologies},
year = {2026},
volume = {14},
number = {6},
pages = {349},
doi = {10.3390/technologies14060349}
}