EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A Detuning-Resilient Wideband Microstrip Antenna for Wearable Biomedical Telemetry Through Progressive Interdigital Slot Integration and Patch Defect Engineering

Prince O Siaw, Ebenezer Adjei iD, Ahmad Aldelemy, John G Buckley iD, Raed Abd-Alhameed iD

Sensors · 2026

DOI: 10.3390/s26196117 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2026. Verfasst von Prince O Siaw und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Digital and Sustainable Futures, University of Bradford, Bradford BD7 1DP, UK. Hauptthemen laut MEDLINE: Telemetry, Wearable Electronic Devices. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council, UK Engineering and Physical Sciences Research Council, Marie Skłodowska-Curie, Research and Innovation Staff Exchange.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents the systematic design, equivalent circuit analysis, and experimental validation of a compact wideband circular microstrip patch antenna for use in biomedical telemetry. The antenna, loaded with interdigital capacitive slots and a defected patch structure (DPS), was fabricated on a Rogers RO3006 substrate (relative permittivity 6.5, loss tangent 0.002, thickness 0.8 mm) with overall dimensions of 25 × 36.96 mm2. A stepwise design evolution through five variants (ANT1–ANT5) illustrates the progressive bandwidth enhancement mechanisms. The final design (ANT5) achieved a simulated −10 dB impedance bandwidth of 1.50 GHz (2.51–4.01 GHz), corresponding to a fractional bandwidth of 46%. The fabricated prototype exhibited a measured −10 dB bandwidth of 1.42 GHz (2.54–3.96 GHz), representing a fractional bandwidth of 43.7%. An equivalent lumped-element circuit model was derived and validated against the full-wave simulation results. The antenna exhibited a simulated peak free-space gain of 3.2 dBi at 3.8 GHz, with free-space radiation efficiency exceeding 90%. Under multi-layer tissue phantom loading, the antenna maintained a robust 910 MHz operational bandwidth with a safe maximum specific absorption rate (SAR) of 0.724 W/kg and an on-body efficiency of 8.4%. The compact size and wide fractional bandwidth provide enhanced resilience against environmental detuning, making this design particularly suitable for robust wearable biomedical telemetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TelemetryWearable Electronic DevicesEquipment DesignWireless TechnologyHumans

Stichwörter der Autor:innen

bandwidth enhancementcircular microstrip patch antennacompact antennadefected patch structuredetuning resilienceequivalent circuit modelinterdigital slotsspecific absorption ratesub-6 GHzwideband antenna

Autor:innen und Institutionen

  • Institute of Digital and Sustainable Futures, University of Bradford, Bradford BD7 1DP, UK.
  • Management, Sciences and Engineering, University of Bradford, Bradford BD7 1DP, UK.
  • Department of Information and Communication Engineering, Alfarqadein University, Basrah 61004, Iraq.

ORCID-Profile: Ebenezer Adjei, John G Buckley, Raed Abd-Alhameed

Förderung

  • Engineering and Physical Sciences Research Council
  • UK Engineering and Physical Sciences Research Council
  • Marie Skłodowska-Curie, Research and Innovation Staff Exchange

Zitieren

APA

Siaw, P. O., Adjei, E., Aldelemy, A., Buckley, J. G., & Abd-Alhameed, R. (2026). A Detuning-Resilient Wideband Microstrip Antenna for Wearable Biomedical Telemetry Through Progressive Interdigital Slot Integration and Patch Defect Engineering. Sensors. https://doi.org/10.3390/s26196117

Vancouver

Siaw PO, Adjei E, Aldelemy A, Buckley JG, Abd-Alhameed R. A Detuning-Resilient Wideband Microstrip Antenna for Wearable Biomedical Telemetry Through Progressive Interdigital Slot Integration and Patch Defect Engineering. Sensors. 2026. doi:10.3390/s26196117

BibTeX
@article{Siaw2026,
  author = {Siaw, Prince O and Adjei, Ebenezer and Aldelemy, Ahmad and Buckley, John G and Abd-Alhameed, Raed},
  title = {A Detuning-Resilient Wideband Microstrip Antenna for Wearable Biomedical Telemetry Through Progressive Interdigital Slot Integration and Patch Defect Engineering},
  journal = {Sensors},
  year = {2026},
  doi = {10.3390/s26196117}
}