EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Quad-Band 1 × 4 Linear MIMO Antenna for Millimeter-Wave, Wearable and Biomedical Telemetry Applications

Rakesh N Tiwari iD, K Geetha Malya, Girigari Nandini, P Baby Nikhitha, Deepti Sharma iD, Prabhakar Singh iD, Pradeep Kumar iD

Sensors · 2024

DOI: 10.3390/s24144427 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Rakesh N Tiwari und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics & Communication Engineering, Madanapalle Institute of Technology & Science, Madanapalle 517325, Andhra Pradesh, India. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Telemetry. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this paper, we present the design of a millimeter-wave 1 × 4 linear MIMO array antenna that operates across multiple resonance frequency bands: 26.28–27.36 GHz, 27.94–28.62 GHz, 32.33–33.08 GHz, and 37.59–39.47 GHz, for mm-wave wearable biomedical telemetry application. The antenna is printed on a flexible substrate with dimensions of 11.0 × 44.0 mm2. Each MIMO antenna element features a modified slot-loaded triangular patch, incorporating ‘cross’-shaped slots in the ground plane to improve impedance matching. The MIMO antenna demonstrates peak gains of 6.12, 8.06, 5.58, and 8.58 dBi at the four resonance frequencies, along with a total radiation efficiency exceeding 75%. The proposed antenna demonstrates excellent diversity metrics, with an ECC < 0.02, DG > 9.97 dB, and CCL below 0.31 bits/sec/Hz, indicating high performance for mm-wave applications. To verify its properties under flexible conditions, a bending analysis was conducted, showing stable S-parameter results with deformation radii of 40 mm (Rx) and 25 mm (Ry). SAR values for the MIMO antenna are calculated at 28.0/38.0 GHz. The average SAR values for 1 gm/10 gm of tissues at 28.0 GHz are found to be 0.0125/0.0079 W/Kg, whereas, at 38.0 GHz, average SAR values are 0.0189/0.0094 W/Kg, respectively. Additionally, to demonstrate the telemetry range of biomedical applications, a link budget analysis at both 28.0 GHz and 38.0 GHz frequencies indicated strong signal strength of 33.69 dB up to 70 m. The fabricated linear MIMO antenna effectively covers the mm-wave 5G spectrum and is suitable for wearable and biomedical applications due to its flexible characteristics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Wearable Electronic DevicesTelemetryHumansWireless TechnologyEquipment Design

Stichwörter der Autor:innen

5G and WBAN applicationsLinear MIMO antennasSARbending analysisdiversity parameterslink budgetmm-wavequad band

Autor:innen und Institutionen

  • Department of Electronics & Communication Engineering, Madanapalle Institute of Technology & Science, Madanapalle 517325, Andhra Pradesh, India.
  • Department of Electronics and Communication Engineering, G. L. Bajaj Institute of Technology and Management, Greater Noida 201306, Uttar Pradesh, India.
  • Physics Division, School of Basic Sciences, Galgotias University, Greater Noida 203201, Uttar Pradesh, India.
  • Discipline of Electrical, Electronic and Computer Engineering, University of KwaZulu-Natal, Durban 4041, South Africa.

ORCID-Profile: Rakesh N Tiwari, Deepti Sharma, Prabhakar Singh, Pradeep Kumar

Zitieren

APA

Tiwari, R. N., Malya, K. G., Nandini, G., Nikhitha, P. B., Sharma, D., Singh, P., & Kumar, P. (2024). Quad-Band 1 × 4 Linear MIMO Antenna for Millimeter-Wave, Wearable and Biomedical Telemetry Applications. Sensors. https://doi.org/10.3390/s24144427

Vancouver

Tiwari RN, Malya KG, Nandini G, Nikhitha PB, Sharma D, Singh P, et al. Quad-Band 1 × 4 Linear MIMO Antenna for Millimeter-Wave, Wearable and Biomedical Telemetry Applications. Sensors. 2024. doi:10.3390/s24144427

BibTeX
@article{Tiwari2024,
  author = {Tiwari, Rakesh N and Malya, K Geetha and Nandini, Girigari and Nikhitha, P Baby and Sharma, Deepti and Singh, Prabhakar and Kumar, Pradeep},
  title = {Quad-Band 1 × 4 Linear MIMO Antenna for Millimeter-Wave, Wearable and Biomedical Telemetry Applications},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s24144427}
}
Kostenlos starten →