Concurrent Validity and Within-Session Reliability of a Wireless Surface Electromyography Device (MR EMG) Compared to a Criterion Measure During the Leg Extension Exercise

Jamie J Ghigiarelli, Adam M Gonzalez, Grace A Valdez iD, Thomas L Johannessen, Admar L Calovini, James M Hackney

Sensors · 14.08.2026 · 26 (16) · 5165

DOI: 10.3390/s26165165 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 14.08.2026. Verfasst von Jamie J Ghigiarelli und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Allied Health and Kinesiology, Hofstra University, Hempstead, NY 11549, USA. Hauptthemen laut MEDLINE: Electromyography, Leg, Wireless Technology, Exercise. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study was to assess within-session reliability and concurrent validity of a wireless surface electromyography device (MR EMG) relative to a wireless criterion system (Delsys Trigno) during dynamic leg extension at varying loading intensities.

Methods

Thirty-one resistance-trained adults performed four repetitions at 30%, 60%, and 90% of their predicted one-repetition maximum (1RM) on the plate-loaded leg extension machine, with EMG amplitude recorded simultaneously by both systems for the vastus lateralis (VL) and vastus medialis obliquus (VMO). Root mean square (RMS) amplitudes were normalized to maximal voluntary isometric contraction and compared across devices using intraclass correlation coefficients (ICC), coefficients of variation (CV%), standard error of measurement, Pearson correlations, and Bland–Altman analyses.

Results

Both devices demonstrated excellent within-session reliability (ICC = 0.92–0.97; CV = 7.8–11.3%). Validity differed by muscle. Across all loads, VMO agreement was good (ICC = 0.82–0.85; r = 0.83–0.86), while VL agreement was moderate (ICC = 0.53–0.57; r = 0.54–0.64). Bland–Altman analyses revealed systematic positive bias across all conditions, with proportional bias present for the VL but absent for the VMO.

Conclusions

MR EMG demonstrated excellent within-session reliability but muscle-dependent concurrent validity, with good VMO agreement and moderate VL agreement with the criterion system.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectromyographyLegWireless TechnologyExerciseHumansAdultReproducibility of ResultsMaleIsometric ContractionFemaleMuscle, SkeletalYoung Adult

Stichwörter der Autor:innen

RMS amplitudeelectromyographyleg extensionreliabilityvalidityvastus lateralisvastus medialis

Autor:innen und Institutionen

  • Department of Allied Health and Kinesiology, Hofstra University, Hempstead, NY 11549, USA.

ORCID-Profile: Grace A Valdez

Zitieren

APA

Ghigiarelli, J. J., Gonzalez, A. M., Valdez, G. A., Johannessen, T. L., Calovini, A. L., & Hackney, J. M. (2026). Concurrent Validity and Within-Session Reliability of a Wireless Surface Electromyography Device (MR EMG) Compared to a Criterion Measure During the Leg Extension Exercise. Sensors, 26(16), 5165. https://doi.org/10.3390/s26165165

Vancouver

Ghigiarelli JJ, Gonzalez AM, Valdez GA, Johannessen TL, Calovini AL, Hackney JM. Concurrent Validity and Within-Session Reliability of a Wireless Surface Electromyography Device (MR EMG) Compared to a Criterion Measure During the Leg Extension Exercise. Sensors. 2026;26(16):5165. doi:10.3390/s26165165

BibTeX
@article{Ghigiarelli2026,
  author = {Ghigiarelli, Jamie J and Gonzalez, Adam M and Valdez, Grace A and Johannessen, Thomas L and Calovini, Admar L and Hackney, James M},
  title = {Concurrent Validity and Within-Session Reliability of a Wireless Surface Electromyography Device (MR EMG) Compared to a Criterion Measure During the Leg Extension Exercise},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {16},
  pages = {5165},
  doi = {10.3390/s26165165}
}
Kostenlos starten →