Wireless versus conventional standardized surface electromyography during clenching: Patient perception and data repeatability

Giacomo Begnoni iD, Guy Willems iD, Sonia Coman iD, Julie De Vlieger iD, Isa den Boer iD, Lies Dekoster iD, Anke Vandeberg iD, Emilie Willems iD, Maria Cadenas de Llano-Pérula iD

European Journal of Oral Sciences · 2025

DOI: 10.1111/eos.70057 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift European Journal of Oral Sciences, 2025. Verfasst von Giacomo Begnoni und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Health Sciences-Orthodontics, KU Leuven and Dentistry, University Hospital Leuven, Leuven, Belgium. Hauptthemen laut MEDLINE: Electromyography, Wireless Technology. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this study was to compare patient perception and data repeatability of conventional and wireless standardized surface electromyography (ssEMG). The electromyographic recordings with both devices were collected at two timepoints (T1 and T2; mean interval 26 days). At T1, participants rated comfort, quickness, and stress on a 10‐point Likert scale, with differences assessed using Student's t ‐test. Data repeatability was analyzed through a linear mixed model (LMM), testing the effects of device type and timepoint on EMG values. Repeatability of ssEMG indices across T1 and T2 was further examined with intraclass correlation coefficients (ICCs). A final sample of 45 participants (37 women, mean age 22.8 ± 1.3 years) was included. Wireless ssEMG was rated significantly more comfortable and quicker. LMM revealed no statistically significant main effects for device (estimate 1.7, 95% CI −3.83 to 7.23) or for timepoint (estimate 0.11, 95% CI −5.41 to 5.64), indicating consistent performance across conditions. Despite moderate‐to‐good within‐ and between‐device reliability (ICC), several indices showed considerable variability across measurements. Concluding, wireless ssEMG is preferred by patients due to their comfort and quickness. ssEMG indices demonstrate good repeatability across devices, but the substantial variability limits their diagnostic utility.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectromyographyWireless TechnologyHumansFemaleReproducibility of ResultsMaleYoung AdultAdult

Stichwörter der Autor:innen

analysisclenchingreproducibilityssEMGsurface electromyography

Autor:innen und Institutionen

  • Department of Oral Health Sciences-Orthodontics, KU Leuven and Dentistry, University Hospital Leuven, Leuven, Belgium.

ORCID-Profile: Giacomo Begnoni, Guy Willems, Sonia Coman, Julie De Vlieger, Isa den Boer, Lies Dekoster, Anke Vandeberg, Emilie Willems, Maria Cadenas de Llano-Pérula

Zitieren

APA

Begnoni, G., Willems, G., Coman, S., De Vlieger, J., den Boer, I., Dekoster, L., Vandeberg, A., Willems, E., & de Llano-Pérula, M. C. (2025). Wireless versus conventional standardized surface electromyography during clenching: Patient perception and data repeatability. European Journal of Oral Sciences. https://doi.org/10.1111/eos.70057

Vancouver

Begnoni G, Willems G, Coman S, De Vlieger J, den Boer I, Dekoster L, et al. Wireless versus conventional standardized surface electromyography during clenching: Patient perception and data repeatability. European Journal of Oral Sciences. 2025. doi:10.1111/eos.70057

BibTeX
@article{Begnoni2025,
  author = {Begnoni, Giacomo and Willems, Guy and Coman, Sonia and De Vlieger, Julie and den Boer, Isa and Dekoster, Lies and Vandeberg, Anke and Willems, Emilie and de Llano-Pérula, Maria Cadenas},
  title = {Wireless versus conventional standardized surface electromyography during clenching: Patient perception and data repeatability},
  journal = {European Journal of Oral Sciences},
  year = {2025},
  doi = {10.1111/eos.70057}
}
Kostenlos starten →