EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Concurrent Validity and Reliability Between Noraxon Ultium™ IMU and Vicon OMC for Lower Limb Gait Assessment Across Variable Walking Speeds

Trung T Le iD, Ha V Vo, Scott C E Brandon iD

Sensors · 16.07.2026 · 26 (14) · 4520

DOI: 10.3390/s26144520 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 16.07.2026. Verfasst von Trung T Le und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, School of Engineering, Mercer University, Macon, GA 31207, USA. Hauptthemen laut MEDLINE: Lower Extremity, Gait, Walking Speed, Walking. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada, Mercer On Mission-Prosthetic Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Wearable inertial measurement units (IMoCAPs) are increasingly used in clinical gait analysis due to their portability and ability to capture data outside laboratory settings; however, validation across operating conditions is essential.

Objective

To evaluate the concurrent validity and reliability of the Noraxon Ultium™ IMU system against a Vicon optical motion capture (OMC) system for lower-limb kinematics during walking across different speeds and time intervals.

Methods

Ten healthy adults performed overground walking at slow, normal, and fast self-selected speeds. Kinematics were recorded simultaneously using both systems. Discrete variables (Max, Min, ROM) were analyzed using three-factor repeated-measures ANOVA (Device × Speed × Time, p < 0.05). Agreement was assessed using Bland–Altman analysis, RMSE, and ICC (3,1), and waveform differences were evaluated using Statistical Parametric Mapping (SPM).

Results

Time effects were minimal across all planes. Sagittal-plane kinematics showed strong agreement, with small biases (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lower ExtremityGaitWalking SpeedWalkingGait AnalysisHumansBiomechanical PhenomenaMotion CaptureReproducibility of ResultsAdultMaleFemaleWearable Electronic Devices

Stichwörter der Autor:innen

Bland–Atman agreementIMoCAPOMCRMSEgait analysislower-limb kinematicsreliabilityvalidationwalking speed

Autor:innen und Institutionen

  • Department of Biomedical Engineering, School of Engineering, Mercer University, Macon, GA 31207, USA.
  • Department of Interdisciplinary Engineering, College of Engineering, University of Guelph, Guelph, ON N1G 2W1, Canada.
  • Department of Orthopedic Surgery, School of Medicine, Mercer University, Macon, GA 31207, USA.

ORCID-Profile: Trung T Le, Scott C E Brandon

Förderung

  • Natural Sciences and Engineering Research Council of Canada
  • Mercer On Mission-Prosthetic Program

Zitieren

APA

Le, T. T., Vo, H. V., & Brandon, S. C. E. (2026). Concurrent Validity and Reliability Between Noraxon Ultium™ IMU and Vicon OMC for Lower Limb Gait Assessment Across Variable Walking Speeds. Sensors, 26(14), 4520. https://doi.org/10.3390/s26144520

Vancouver

Le TT, Vo HV, Brandon SCE. Concurrent Validity and Reliability Between Noraxon Ultium™ IMU and Vicon OMC for Lower Limb Gait Assessment Across Variable Walking Speeds. Sensors. 2026;26(14):4520. doi:10.3390/s26144520

BibTeX
@article{Le2026,
  author = {Le, Trung T and Vo, Ha V and Brandon, Scott C E},
  title = {Concurrent Validity and Reliability Between Noraxon Ultium™ IMU and Vicon OMC for Lower Limb Gait Assessment Across Variable Walking Speeds},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {14},
  pages = {4520},
  doi = {10.3390/s26144520}
}
Kostenlos starten →