From PPG to Blood Pressure at the Edge: Quantization-Aware Architecture Selection and On-MCU Validation
Sensors · 25.04.2026 · 26 (9) · 2674
DOI: 10.3390/s26092674 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 25.04.2026. Verfasst von Elisabetta Leogrande und 2 weiteren Autor:innen. Beteiligte Einrichtung: Micro Nano Sensor Group, Polytechnic University of Bari, Via Edoardo Orabona 4, 70126 Bari, Italy. Hauptthemen laut MEDLINE: Photoplethysmography, Blood Pressure, Blood Pressure Determination. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Blood pressure is a central marker of cardiovascular risk, but continuous monitoring remains difficult because cuff-based measurements are intermittent and uncomfortable. Photoplethysmography (PPG) is already ubiquitous in wearables and can, in principle, enable cuffless blood pressure estimation from a single optical signal. However, many deep learning approaches that perform well in floating-point are impractical for microcontroller-class devices, where memory budgets, latency, and integer-only arithmetic constrain what can be deployed. A key open question is which neural architectures retain accuracy after full-integer quantization, rather than only under desktop inference. Here, we show an end-to-end, microcontroller-oriented evaluation framework that benchmarks multiple 1D convolutional models for cuffless systolic and diastolic pressure estimation from single-channel PPG, jointly optimizing estimation error, model footprint, and quantization robustness. We find that floating-point accuracy alone is a poor predictor of deployability: some lightweight CNNs exhibit substantial performance drift after INT8 conversion, whereas a compact residual 1D CNN preserves its predictions with near-identical error statistics after integer quantization. We then deploy the selected integer-only model on an STM32N6 microcontroller using an industrial toolchain and confirm that on-device inference maintains low bias and limited error dispersion while meeting real-time constraints for continuous operation. These results highlight architecture-dependent quantization stability as a critical design dimension for sensor-edge intelligence and support the feasibility of fully on-device cuffless blood pressure monitoring without multimodal sensing or cloud processing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Micro Nano Sensor Group, Polytechnic University of Bari, Via Edoardo Orabona 4, 70126 Bari, Italy.
ORCID-Profile: Elisabetta Leogrande, Emanuele De Luca, Francesco Dell'Olio
Zitieren
APA
Leogrande, E., Luca, E., & Dell'Olio, F. (2026). From PPG to Blood Pressure at the Edge: Quantization-Aware Architecture Selection and On-MCU Validation. Sensors, 26(9), 2674. https://doi.org/10.3390/s26092674
Vancouver
Leogrande E, Luca E, Dell'Olio F. From PPG to Blood Pressure at the Edge: Quantization-Aware Architecture Selection and On-MCU Validation. Sensors. 2026;26(9):2674. doi:10.3390/s26092674
BibTeX
@article{Leogrande2026,
author = {Leogrande, Elisabetta and Luca, Emanuele De and Dell'Olio, Francesco},
title = {From PPG to Blood Pressure at the Edge: Quantization-Aware Architecture Selection and On-MCU Validation},
journal = {Sensors},
year = {2026},
volume = {26},
number = {9},
pages = {2674},
doi = {10.3390/s26092674}
}