Real-Time ECG Artifact Removal for Adaptive Deep Brain Stimulation: A Comparative Study
Sensors · 25.04.2026 · 26 (9) · 2673
DOI: 10.3390/s26092673 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Sensors, online erschienen am 25.04.2026. Verfasst von Lucrezia Silvi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Biomedical Engineering Research to Advance and Innovate Translational Neuroscience (BRAIN Unit), Department of Neuroscience, University of Padova, 35…. Hauptthemen laut MEDLINE: Deep Brain Stimulation, Artifacts, Electrocardiography, Parkinson Disease. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: PRIN.
Abstract (Originalsprache)
Background
Adaptive deep brain stimulation (aDBS) for Parkinson’s disease (PD) relies on accurate detection of beta oscillatory activity. However, electrocardiographic (ECG) artifacts frequently contaminate local field potentials (LFPs), compromising control algorithms. While offline cleaning methods exist, their feasibility for real-time operation within the strict timing constraints of current sensing-enabled devices remains unknown.
Methods
We evaluated four ECG removal algorithms, template subtraction (TS), singular value decomposition (SVD), extended SVD (eSVD), and the Perceive toolbox (PR), on simulated datasets (contaminated at −30 to +20 dB) and clinical recordings from 20 PD patients. Algorithms were assessed for artifact removal quality (beta power preservation, signal-to-noise ratio) and real-time feasibility (99th percentile processing latency—P99 < 50 ms).
Results
Only TS and standard SVD met the real-time feasibility threshold, with TS achieving superior timing consistency (P99 ≈ 10 ms). eSVD and PR proved incompatible with closed-loop requirements (P99 > 90 ms). While eSVD yielded the highest artifact suppression at extreme contamination, it suffered from poor signal preservation at moderate levels. TS demonstrated the best balance, maintaining beta power accuracy within ±12% across clinically relevant contamination levels.
Conclusions
TS is the recommended method for real-time aDBS applications, offering a safety-critical balance of computational efficiency and biomarker fidelity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Biomedical Engineering Research to Advance and Innovate Translational Neuroscience (BRAIN Unit), Department of Neuroscience, University of Padova, 35122 Padua, Italy.
- Padova Neuroscience Center (PNC), University of Padova, 35122 Padua, Italy.
- Department of Neuroscience, University of Padova, 35122 Padua, Italy.
- Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS), San Camillo Hospital, 30126 Venice, Italy.
- Institute of Cognitive Sciences and Technologies (ISTC), National Research Council (CNR), 00185 Rome, Italy.
- Center of Human Brain Health, School of Psychology, University of Birmingham, Birmingham B15 2TT, UK.
ORCID-Profile: Valentina D'Onofrio, Angelo Antonini, Andrea Guerra, Camillo Porcaro
Förderung
- PRIN
Zitieren
APA
Silvi, L., D'Onofrio, V., Cauzzo, S., Antonini, A., Guerra, A., & Porcaro, C. (2026). Real-Time ECG Artifact Removal for Adaptive Deep Brain Stimulation: A Comparative Study. Sensors, 26(9), 2673. https://doi.org/10.3390/s26092673
Vancouver
Silvi L, D'Onofrio V, Cauzzo S, Antonini A, Guerra A, Porcaro C. Real-Time ECG Artifact Removal for Adaptive Deep Brain Stimulation: A Comparative Study. Sensors. 2026;26(9):2673. doi:10.3390/s26092673
BibTeX
@article{Silvi2026,
author = {Silvi, Lucrezia and D'Onofrio, Valentina and Cauzzo, Simone and Antonini, Angelo and Guerra, Andrea and Porcaro, Camillo},
title = {Real-Time ECG Artifact Removal for Adaptive Deep Brain Stimulation: A Comparative Study},
journal = {Sensors},
year = {2026},
volume = {26},
number = {9},
pages = {2673},
doi = {10.3390/s26092673}
}
EUVIMED