Validation and Reproducibility of an App for Continuous Measurement as an Assessment Tool for Idiopathic Scoliosis
Sensors · 27.03.2026 · 26 (7) · 2099
DOI: 10.3390/s26072099 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 27.03.2026. Verfasst von Isis Juliene Rodrigues Leite Navarro und 7 weiteren Autor:innen. Beteiligte Einrichtung: ISICO (Italian Scientific Spine Institute), 20141 Milan, Italy. Hauptthemen laut MEDLINE: Scoliosis, Mobile Applications. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
(1)
Background
Idiopathic scoliosis is a three-dimensional deformity, yet clinical and research decision-making still relies largely on radiographic Cobb angle measurements. As a radiation-free alternative, clinical assessment of transverse and sagittal plane deformities has gained importance. This study evaluated the concurrent validity and intra- and interrater reproducibility of continuous measurements of rib hump, thoracic kyphosis, and lumbar lordosis obtained using a smartphone application in adolescents with spinal deformities. (2)
Methods
Adolescents aged 10–17 years with scoliosis (>10° Cobb) or hyperkyphosis (>50° Cobb) were recruited. Continuous measurements of angle of trunk rotation (ATR) during the Adams forward bend test and in standing position, as well as sagittal profile, were collected using the ISICO app mounted on a standardized plastic tool. Concurrent validity was assessed against a scoliometer using Spearman correlation, root mean square error, and Bland–Altman analysis, while reproducibility was evaluated using intraclass correlation coefficients, standard error of measurement, and minimal detectable change. (3)
Results
Thirty-two adolescents were included for validation and intrarater analyses and 34 for interrater analyses. ATR measured during the Adams test showed very high correlation with the scoliometer and minimal bias, while standing ATR showed moderate correlation. Reliability was excellent for rib hump during forward bending and moderate for sagittal parameters, with the lowest values observed for lumbar lordosis. (4)
Conclusions
These findings support the clinical use of continuous app-based ATR assessment and suggest that sagittal measurements may be useful with appropriate examiner training.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- ISICO (Italian Scientific Spine Institute), 20141 Milan, Italy.
- Department of Physical Medicine and Rehabilitation, Lariboisière-Fernand Widal Hospital, AP-HP, Université Paris Cité, 75010 Paris, France.
- Epidemiology of Ageing and Neurodegenerative Diseases (EpiAgeing), Inserm U1153, Université Paris Cité, 75010 Paris, France.
- Research and Development Unit, Parc Sanitari Sant Joan de Déu, CIBERSAM, ISCIII, Dr. Antoni Pujadas, 42, Sant Boi de Llobregat, 08830 Barcelona, Spain.
- University of Milan, 20122 Milan, Italy.
- IRCCS Galeazzi S.Ambrogio Hospital, 20161 Milan, Italy.
ORCID-Profile: Isis Juliene Rodrigues Leite Navarro, Francesco Dulio, Stefano Negrini, Fabio Zaina, Alessandra Negrini
Zitieren
APA
Navarro, I. J. R. L., Jacob, L., Masetto, K., Dulio, F., Negrini, A., Negrini, S., Zaina, F., & Negrini, A. (2026). Validation and Reproducibility of an App for Continuous Measurement as an Assessment Tool for Idiopathic Scoliosis. Sensors, 26(7), 2099. https://doi.org/10.3390/s26072099
Vancouver
Navarro IJRL, Jacob L, Masetto K, Dulio F, Negrini A, Negrini S, et al. Validation and Reproducibility of an App for Continuous Measurement as an Assessment Tool for Idiopathic Scoliosis. Sensors. 2026;26(7):2099. doi:10.3390/s26072099
BibTeX
@article{Navarro2026,
author = {Navarro, Isis Juliene Rodrigues Leite and Jacob, Louis and Masetto, Kevin and Dulio, Francesco and Negrini, Andrea and Negrini, Stefano and Zaina, Fabio and Negrini, Alessandra},
title = {Validation and Reproducibility of an App for Continuous Measurement as an Assessment Tool for Idiopathic Scoliosis},
journal = {Sensors},
year = {2026},
volume = {26},
number = {7},
pages = {2099},
doi = {10.3390/s26072099}
}
EUVIMED