An Analytical Model of Motion Artifacts in a Measured Arterial Pulse Signal—Part I: Accelerometers and PPG Sensors
Sensors · 12.09.2025 · 25 (18) · 5710
DOI: 10.3390/s25185710 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 12.09.2025. Verfasst von Md Mahfuzur Rahman und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical and Aerospace Engineering, Old Dominion University, Norfolk, VA 23529, USA. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This paper, the first of two parts, presents an analytical model of motion artifacts (MAs) in measured pulse signals by accelerometers and photoplethysmography (PPG) sensors. As the transmission path from the true pulse signal in an artery to the sensor output (measured pulse signal), the tissue–contact–sensor (TCS) stack is modeled as a 1DOF (degree-of-freedom) system. MAs cause baseline drift of the mass and simultaneously time-varying system parameters (TVSPs) of the TCS stack. With arterial wall displacement and pulsatile pressure serving separately as the true pulse signal, an analytical model is developed to mathematically relate baseline drift and TVSP to a measured pulse signal. With assumed values of baseline drift and TVSPs, the numerical calculation is conducted in MATLAB. While baseline drift is low-frequency additive noise and can greatly swing a measured pulse signal, TVSP generates relatively small, abrupt distortion (e.g., 1% variation in heart rate and
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical and Aerospace Engineering, Old Dominion University, Norfolk, VA 23529, USA.
ORCID-Profile: Md Mahfuzur Rahman, Subodh Toraskar, Zhili Hao
Zitieren
APA
Rahman, M. M., Toraskar, S., Hasan, M., & Hao, Z. (2025). An Analytical Model of Motion Artifacts in a Measured Arterial Pulse Signal—Part I: Accelerometers and PPG Sensors. Sensors, 25(18), 5710. https://doi.org/10.3390/s25185710
Vancouver
Rahman MM, Toraskar S, Hasan M, Hao Z. An Analytical Model of Motion Artifacts in a Measured Arterial Pulse Signal—Part I: Accelerometers and PPG Sensors. Sensors. 2025;25(18):5710. doi:10.3390/s25185710
BibTeX
@article{Rahman2025,
author = {Rahman, Md Mahfuzur and Toraskar, Subodh and Hasan, Mamun and Hao, Zhili},
title = {An Analytical Model of Motion Artifacts in a Measured Arterial Pulse Signal—Part I: Accelerometers and PPG Sensors},
journal = {Sensors},
year = {2025},
volume = {25},
number = {18},
pages = {5710},
doi = {10.3390/s25185710}
}
EUVIMED