EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Design and Validation of Low-Cost, Portable Impedance Analyzer System for Biopotential Electrode Evaluation and Skin/Electrode Impedance Measurement

Jaydeep Panchal iD, Moon Inder Singh iD, Mandeep Singh iD, Karmjit Singh Sandha

Sensors · 12.06.2025 · 25 (12) · 3688

DOI: 10.3390/s25123688 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 12.06.2025. Verfasst von Jaydeep Panchal und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Instrumentation Engineering, Thapar Institute of Engineering and Technology, Bhadson Rd, Patiala 147004, Punjab, India. Hauptthemen laut MEDLINE: Electric Impedance, Skin, Biosensing Techniques. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Thapar Institute of Engineering and Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents a novel, low-cost, portable impedance analyzer system designed for biopotential electrode evaluation and skin/electrode impedance measurement, critical for enhancing bioelectrical signal quality in healthcare applications. In contrast with conventional systems that depend on external PCs or host devices for data acquisition, visualization, and analysis, this design integrates all functionalities into a single, compact platform powered by the Analog Devices AD5933 impedance converter and a Raspberry Pi 4. The design incorporates custom analog circuitry to extend the measurement range from 10 Hz to 100 kHz and supports a wide impedance spectrum through switchable feedback resistors. Validated against a benchtop impedance analyzer, the system demonstrates high accuracy with normalized root-mean-square errors (NRMSEs) of 1.41% and 3.77% for the impedance magnitude and phase of passive components, respectively, and 1.43% and 1.29% for the biopotential electrode evaluation and skin/electrode impedance measurement. This cost-effective solution, with a total cost of USD 159, addresses the accessibility challenges faced by smaller research labs and healthcare facilities, offering a compact, low-power platform for reliable impedance analysis in biomedical applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Electric ImpedanceSkinBiosensing TechniquesElectrodesHumansEquipment Design

Stichwörter der Autor:innen

AD5933biopotential electrodeimpedance analyzerimpedance measurementportableskin/electrode impedance

Autor:innen und Institutionen

  • Department of Electrical and Instrumentation Engineering, Thapar Institute of Engineering and Technology, Bhadson Rd, Patiala 147004, Punjab, India.

ORCID-Profile: Jaydeep Panchal, Moon Inder Singh, Mandeep Singh

Förderung

  • Thapar Institute of Engineering and Technology

Zitieren

APA

Panchal, J., Singh, M. I., Singh, M., & Sandha, K. S. (2025). Design and Validation of Low-Cost, Portable Impedance Analyzer System for Biopotential Electrode Evaluation and Skin/Electrode Impedance Measurement. Sensors, 25(12), 3688. https://doi.org/10.3390/s25123688

Vancouver

Panchal J, Singh MI, Singh M, Sandha KS. Design and Validation of Low-Cost, Portable Impedance Analyzer System for Biopotential Electrode Evaluation and Skin/Electrode Impedance Measurement. Sensors. 2025;25(12):3688. doi:10.3390/s25123688

BibTeX
@article{Panchal2025,
  author = {Panchal, Jaydeep and Singh, Moon Inder and Singh, Mandeep and Sandha, Karmjit Singh},
  title = {Design and Validation of Low-Cost, Portable Impedance Analyzer System for Biopotential Electrode Evaluation and Skin/Electrode Impedance Measurement},
  journal = {Sensors},
  year = {2025},
  volume = {25},
  number = {12},
  pages = {3688},
  doi = {10.3390/s25123688}
}
Kostenlos starten →