Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study
Sensors · 2024
DOI: 10.3390/s24216883 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Sensors, 2024. Verfasst von Djamel Dabli und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging, IMAGINE UR UM 103, Montpellier University, Nimes University Hospital, Bd Prof Robert Debré, CEDEX 9, 30029 Nîmes, Fran…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Imaging, Three-Dimensional, Spine, Tomography, X-Ray Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
An image-quality CT phantom was scanned with three different 3D X-ray imaging guidance devices in the operating theatre: O-Arm, Loop-X, and Airo TruCT. Default acquisition and reconstruction parameters for lumbar spine procedures were used on each device. The tube current was set to a dose level of around 27 mGy. A task-based image quality assessment was performed by calculating the noise power spectrum (NPS) and task transfer function (TTF). A detectability index (d’) was calculated for three simulated bone lesions. The noise magnitude of the O-Arm was higher than the Airo TruCT, and the Loop-X had higher noise than the Airo TruCT. The highest average NPS frequency was for the O-Arm images, and the lowest was for the Loop-X. The TTFs at 50% values were similar for the Airo TruCT and Loop-X devices. Compared to Airo TruCT, the TTF at 50% value increased with the O-Arm by 53.12% and 41.20% for the Teflon and Delrin inserts, respectively. Compared to Airo TruCT, the d’ value was lower with Loop-X by −26.73%, −27.02%, and −23.95% for lytic lesions, sclerotic lesions, and high-density bone, respectively. Each 3D-imaging spine surgery guidance device has its own strengths and weaknesses in terms of image quality. Cone-beam CT systems apparently offer the best compromise between noise and spatial resolution for spine surgery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Imaging, IMAGINE UR UM 103, Montpellier University, Nimes University Hospital, Bd Prof Robert Debré, CEDEX 9, 30029 Nîmes, France.
- Medical Physics and Radiation Protection Department, APHP Lariboisière University Hospital, 75010 Paris, France.
- Department of Radiology, Georges Pompidou European Hospital, Paris Cité University, APHP, 75015 Paris, France.
- Department of Radiodiagnostics, CHU Angers, 4 Rue Larrey, 49933 Angers, France.
ORCID-Profile: Djamel Dabli, Isabelle Fitton, Claire Van Ngoc Ty, Joël Greffier, Lama Hadid-Beurrier
Zitieren
APA
Dabli, D., Salvat, C., Fitton, I., Van Ngoc Ty, C., Palanchon, P., Beregi, J. P., Greffier, J., & Hadid-Beurrier, L. (2024). Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study. Sensors. https://doi.org/10.3390/s24216883
Vancouver
Dabli D, Salvat C, Fitton I, Van Ngoc Ty C, Palanchon P, Beregi JP, et al. Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study. Sensors. 2024. doi:10.3390/s24216883
BibTeX
@article{Dabli2024,
author = {Dabli, Djamel and Salvat, Cécile and Fitton, Isabelle and Van Ngoc Ty, Claire and Palanchon, Peggy and Beregi, Jean-Paul and Greffier, Joël and Hadid-Beurrier, Lama},
title = {Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24216883}
}
EUVIMED