Efficacy of Wearable Single-Lead ECG Monitoring during Exercise Stress Testing: A Comparative Study
Sensors · 2024
DOI: 10.3390/s24196394 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Sensors, 2024. Verfasst von Hyo-In Choi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Internal Medicine, Sungkyunkwan University School of Medicine, Kangbuk Samsung Hospital, Seoul 03181, Republic …. Hauptthemen laut MEDLINE: Heart Rate, Wearable Electronic Devices, Exercise Test, Electrocardiography. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Seers Co., Ltd..
Abstract (Originalsprache)
Background and Objectives: Few comparative studies have evaluated wearable single-lead electrocardiogram (ECG) devices and standard multi-lead ECG devices during exercise testing. This study aimed to validate the accuracy of a wearable single-lead ECG monitor for recording heart rate (HR) metrics during graded exercise tests (GXTs).
Methods
A cohort of 50 patients at a tertiary hospital underwent GXT while simultaneously being equipped with wearable single- and conventional multi-lead ECGs. The concordance between these modalities was quantified using the intraclass correlation coefficient and Bland–Altman plot analysis.
Results
The minimum and average HR readings between the devices were generally consistent. Parameters such as ventricular ectopic beats and supraventricular ectopic beats showed strong agreement. However, the agreement for the Total QRS and Maximum RR was not sufficient. HR measurements across different stages of the exercise test showed sufficient agreement. Although not statistically significant, the standard multi-lead ECG devices exhibited higher noise levels compared to the wearable single-lead ECG devices.
Conclusions
Wearable single-lead ECG devices can reliably monitor HR and detect abnormal beats across a spectrum of exercise intensities, offering a viable alternative to traditional multi-lead systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiology, Department of Internal Medicine, Sungkyunkwan University School of Medicine, Kangbuk Samsung Hospital, Seoul 03181, Republic of Korea.
- Seers Technology Co., Ltd., Seongnam-si 13558, Republic of Korea.
ORCID-Profile: Hyo-In Choi, Seung Jae Lee, Jong Doo Choi, GyungChul Kim, Young-Shin Lee, Jong-Young Lee
Förderung
- Seers Co., Ltd.
Zitieren
APA
Choi, H. I., Lee, S. J., Choi, J. D., Kim, G., Lee, Y. S., & Lee, J. Y. (2024). Efficacy of Wearable Single-Lead ECG Monitoring during Exercise Stress Testing: A Comparative Study. Sensors. https://doi.org/10.3390/s24196394
Vancouver
Choi HI, Lee SJ, Choi JD, Kim G, Lee YS, Lee JY. Efficacy of Wearable Single-Lead ECG Monitoring during Exercise Stress Testing: A Comparative Study. Sensors. 2024. doi:10.3390/s24196394
BibTeX
@article{Choi2024,
author = {Choi, Hyo-In and Lee, Seung Jae and Choi, Jong Doo and Kim, GyungChul and Lee, Young-Shin and Lee, Jong-Young},
title = {Efficacy of Wearable Single-Lead ECG Monitoring during Exercise Stress Testing: A Comparative Study},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24196394}
}
EUVIMED