EUVIMEDEUVIMED entdecken
VergleichsstudieValidierungsstudieOpen Access · CC BY 4.0

Validation of the Polar Fitness Test for Estimation of Maximal Oxygen Consumption at Rest in Medically Supervised Exercise Training: Comparison with CPET and the 6-Minute Walk Test

Michael Neudorfer, Lukas Ötzlinger, Devender Kumar iD, Josef Niebauer, Jan David Smeddinck iD, Mahdi Sareban, Gunnar Treff iD

Sensors · 2025

DOI: 10.3390/s25185649 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Sensors, 2025. Verfasst von Michael Neudorfer und 6 weiteren Autor:innen. Beteiligte Einrichtung: University Institute of Sports Medicine, Prevention and Rehabilitation, University Hospital Salzburg, Paracelsus Medical University, 5020 Salzburg, A…. Hauptthemen laut MEDLINE: Oxygen Consumption, Walk Test, Exercise Test. Die Arbeit nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The Polar Fitness Test (PFT) estimates maximal oxygen consumption (V̇O2max) under resting conditions using heart rate data from the manufacturer’s wearable devices. We aimed to validate the PFT in a population with cardiovascular risk factors and to compare its results with five established equations predicting V̇O2max based on the 6-min walk test (6MWT). Twenty-four participants (9 female; age 57.4 ± 10.2 years) undergoing medically supervised exercise training—including seven individuals on heart rate-limiting medication—completed the PFT, 6MWT, and cardiopulmonary exercise testing (CPET), which served as the criterion V̇O2max measurement. The PFT showed a mean absolute percent-age error (MAPE) of 13.7%, an intraclass correlation coefficient (ICC) of 0.743, a mean bias of −1.0 mL/min/kg, and limits of agreement (LoA) of ±11.4 mL/min/kg compared to CPET. Among the 6MWT-based equations, only the Porcari equation demonstrated similar performance (MAPE 12.6%, ICC 0.725, mean bias 0.2 mL/min/kg, LoA ± 9.7 mL/min/kg), while other equations showed larger errors and systematic deviations. Our data indicate that the PFT may present an easily accessible option to estimate V̇O2max on population level when exercise-based testing is not feasible. However, its variability limits use for individual clinical decisions, reaffirming the relevance of CPET for accurate assessment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Oxygen ConsumptionWalk TestExercise TestHumansFemaleMiddle AgedMaleHeart RateAgedAdultWearable Electronic DevicesExercise TherapyRestExercise

Stichwörter der Autor:innen

cardiac rehabilitationmHealthphotoplethysmographypreventiontelemedicinewearables

Autor:innen und Institutionen

  • University Institute of Sports Medicine, Prevention and Rehabilitation, University Hospital Salzburg, Paracelsus Medical University, 5020 Salzburg, Austria.
  • Ludwig Boltzmann Institute for Digital Health and Prevention, 5020 Salzburg, Austria.
  • Institute of Molecular Sports and Rehabilitation Medicine, Paracelsus Medical University, 5020 Salzburg, Austria.

ORCID-Profile: Devender Kumar, Jan David Smeddinck, Gunnar Treff

Zitieren

APA

Neudorfer, M., Ötzlinger, L., Kumar, D., Niebauer, J., Smeddinck, J. D., Sareban, M., & Treff, G. (2025). Validation of the Polar Fitness Test for Estimation of Maximal Oxygen Consumption at Rest in Medically Supervised Exercise Training: Comparison with CPET and the 6-Minute Walk Test. Sensors. https://doi.org/10.3390/s25185649

Vancouver

Neudorfer M, Ötzlinger L, Kumar D, Niebauer J, Smeddinck JD, Sareban M, et al. Validation of the Polar Fitness Test for Estimation of Maximal Oxygen Consumption at Rest in Medically Supervised Exercise Training: Comparison with CPET and the 6-Minute Walk Test. Sensors. 2025. doi:10.3390/s25185649

BibTeX
@article{Neudorfer2025,
  author = {Neudorfer, Michael and Ötzlinger, Lukas and Kumar, Devender and Niebauer, Josef and Smeddinck, Jan David and Sareban, Mahdi and Treff, Gunnar},
  title = {Validation of the Polar Fitness Test for Estimation of Maximal Oxygen Consumption at Rest in Medically Supervised Exercise Training: Comparison with CPET and the 6-Minute Walk Test},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25185649}
}