Paper-Based Microfluidic Analytical Device Patterned by Label Printer for Point-of-Care Blood Glucose and Hematocrit Detection Using 3D-Printed Smartphone Cassette

Zong-Xiao Cai, Ming-Zhang Jiang, Ya-Ju Chuang, Ju-Nan Kuo

Sensors · 2024-07-24 · 24 (15) · 4792

DOI: 10.3390/s24154792 ↗

At a glance

Journal article in Sensors, published online on 2024-07-24. Written by Zong-Xiao Cai and 3 co-authors. Institution: Department of Automation Engineering, National Formosa University, No. 64, Wenhua Rd., Huwei 63201, Yunlin, Taiwan. Major MEDLINE topics: Printing, Three-Dimensional, Smartphone, Blood Glucose, Point-of-Care Systems. Cited 12 times so far (Crossref); 54 references. The full text is freely available (open access, CC BY 4.0). Funded by: National Science and Technology Council of Taiwan, National Taiwan University Hospital Yunlin Branch.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

This study presents a portable, low-cost, point-of-care (POC) system for the simultaneous detection of blood glucose and hematocrit. The system consists of a disposable origami microfluidic paper-based analytical device (μPAD) for plasma separation, filtration, and reaction functions and a 3D-printed cassette for hematocrit and blood glucose detection using a smartphone. The origami μPAD is patterned using a cost-effective label printing technique instead of the conventional wax printing method. The 3D-printed cassette incorporates an array of LED lights, which mitigates the effects of intensity variations in the ambient light and hence improves the accuracy of the blood glucose and hematocrit concentration measurements. The hematocrit concentration is determined quantitatively by measuring the distance of plasma wicking along the upper layer of the origami μPAD, which is pretreated with sodium chloride and Tween 20 to induce dehydration and aggregation of the red blood cells. The filtered plasma also penetrates to the lower layer of the origami μPAD, where it reacts with embedded colorimetric assay reagents to produce a yellowish-brown complex. A color image of the reaction complex is captured using a smartphone inserted into the 3D-printed cassette. The image is analyzed using self-written RGB software to quantify the blood glucose concentration. The calibration results indicate that the proposed detection platform provides an accurate assessment of the blood glucose level over the range of 45–630 mg/dL (R2 = 0.9958). The practical feasibility of the proposed platform is demonstrated by measuring the blood glucose and hematocrit concentrations in 13 human whole blood samples. Taking the measurements obtained from commercial glucose and hematocrit meters as a benchmark, the proposed system has a differential of no more than 6.4% for blood glucose detection and 9.1% for hematocrit detection. Overall, the results confirm that the proposed μPAD is a promising solution for cost-effective and reliable POC health monitoring.

Subjects

MeSH terms (MEDLINE)

Printing, Three-DimensionalSmartphoneBlood GlucosePoint-of-Care SystemsPaperHematocritHumansColorimetryLab-On-A-Chip DevicesMicrofluidic Analytical Techniques

Author keywords

glucosehematocritlabel printingmicrofluidicpaper-based devicepoint-of-caresmartphone

Substances

Blood Glucose

Authors and institutions

  • Department of Automation Engineering, National Formosa University, No. 64, Wenhua Rd., Huwei 63201, Yunlin, Taiwan.
  • Department of Laboratory Medicine, National Taiwan University Hospital Yunlin Branch, No. 579, Sec. 2, Yunlin Rd., Douliu 640203, Yunlin, Taiwan.

Funding

  • National Science and Technology Council of Taiwan
  • National Taiwan University Hospital Yunlin Branch

Cite

APA

Cai, Z. X., Jiang, M. Z., Chuang, Y. J., & Kuo, J. N. (2024). Paper-Based Microfluidic Analytical Device Patterned by Label Printer for Point-of-Care Blood Glucose and Hematocrit Detection Using 3D-Printed Smartphone Cassette. Sensors, 24(15), 4792. https://doi.org/10.3390/s24154792

Vancouver

Cai ZX, Jiang MZ, Chuang YJ, Kuo JN. Paper-Based Microfluidic Analytical Device Patterned by Label Printer for Point-of-Care Blood Glucose and Hematocrit Detection Using 3D-Printed Smartphone Cassette. Sensors. 2024;24(15):4792. doi:10.3390/s24154792

BibTeX
@article{Cai2024,
  author = {Cai, Zong-Xiao and Jiang, Ming-Zhang and Chuang, Ya-Ju and Kuo, Ju-Nan},
  title = {Paper-Based Microfluidic Analytical Device Patterned by Label Printer for Point-of-Care Blood Glucose and Hematocrit Detection Using 3D-Printed Smartphone Cassette},
  journal = {Sensors},
  year = {2024},
  volume = {24},
  number = {15},
  pages = {4792},
  doi = {10.3390/s24154792}
}
Start for free →