NIR-Based Electronic Platform for Glucose Monitoring for the Prevention and Control of Diabetes Mellitus
Sensors · 2024
DOI: 10.3390/s24134190 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von William Oñate und 5 weiteren Autor:innen. Beteiligte Einrichtung: Carrera de Electrónica y Automatización, Universidad Politecnica Salesiana (UPS), Quito 170146, Ecuador. Hauptthemen laut MEDLINE: Blood Glucose, Blood Glucose Self-Monitoring, Diabetes Mellitus. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universidad Técnica de Ambato, Universidad Tecnica de Ambato, REDU.
Themen: Diabetes mellitus
Abstract (Originalsprache)
The glucose level in the blood is measured through invasive methods, causing discomfort in the patient, loss of sensitivity in the area where the sample is obtained, and healing problems. This article deals with the design, implementation, and evaluation of a device with an ESP-WROOM-32D microcontroller with the application of near-infrared photospectroscopy technology that uses a diode array that transmits between 830 nm and 940 nm to measure glucose levels in the blood. In addition, the system provides a webpage for the monitoring and control of diabetes mellitus for each patient; the webpage is hosted on a local Linux server with a MySQL database. The tests are conducted on 120 people with an age range of 35 to 85 years; each person undergoes two sample collections with the traditional method and two with the non-invasive method. The developed device complies with the ranges established by the American Diabetes Association: presenting a measurement error margin of close to 3% in relation to traditional blood glucose measurement devices. The purpose of the study is to design and evaluate a device that uses non-invasive technology to measure blood glucose levels. This involves constructing a non-invasive glucometer prototype that is then evaluated in a group of participants with diabetes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Carrera de Electrónica y Automatización, Universidad Politecnica Salesiana (UPS), Quito 170146, Ecuador.
- Faculty of Systems, Electronics and Industrial Engineering, Universidad Tecnica de Ambato (UTA), Ambato 180206, Ecuador.
- Departamento de Ingeniería de Sistemas y Automática, University of the Basque Country, Euskal Herriko Unibertsitatea/Universidad del País Vasco, 48013 Bilbao, Spain.
ORCID-Profile: William Oñate, Edwin Ramos-Zurita, Juan-Pablo Pallo, Santiago Manzano, Paulina Ayala, Marcelo V Garcia
Förderung
- Universidad Técnica de Ambato
- Universidad Tecnica de Ambato
- REDU
Zitieren
APA
Oñate, W., Ramos-Zurita, E., Pallo, J. P., Manzano, S., Ayala, P., & Garcia, M. V. (2024). NIR-Based Electronic Platform for Glucose Monitoring for the Prevention and Control of Diabetes Mellitus. Sensors. https://doi.org/10.3390/s24134190
Vancouver
Oñate W, Ramos-Zurita E, Pallo JP, Manzano S, Ayala P, Garcia MV. NIR-Based Electronic Platform for Glucose Monitoring for the Prevention and Control of Diabetes Mellitus. Sensors. 2024. doi:10.3390/s24134190
BibTeX
@article{Oate2024,
author = {Oñate, William and Ramos-Zurita, Edwin and Pallo, Juan-Pablo and Manzano, Santiago and Ayala, Paulina and Garcia, Marcelo V},
title = {NIR-Based Electronic Platform for Glucose Monitoring for the Prevention and Control of Diabetes Mellitus},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24134190}
}
EUVIMED