The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors

Tomoya Ishida iD, Mina Samukawa iD

Sensors · 2024

DOI: 10.3390/s24072092 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Tomoya Ishida und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Health Sciences, Hokkaido University, North 12, West 5, Kita-ku, Sapporo 060-0812, Japan. Hauptthemen laut MEDLINE: Optical Devices, Wearable Electronic Devices. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nitto (Japan), Nitto Denko Corporation, Tokyo, Japan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Frontal and axial knee motion can affect the accuracy of the knee extension/flexion motion measurement using a wearable goniometer. The purpose of this study was to test the hypothesis that calibrating the goniometer on an individual’s body would reduce errors in knee flexion angle during gait, compared to bench calibration. Ten young adults (23.2 ± 1.3 years) were enrolled. Knee flexion angles during gait were simultaneously assessed using a wearable goniometer sensor and an optical three-dimensional motion analysis system, and the absolute error (AE) between the two methods was calculated. The mean AE across a gait cycle was 2.4° (0.5°) for the on-body calibration, and the AE was acceptable (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Optical DevicesWearable Electronic DevicesYoung AdultHumansCalibrationKnee JointGait

Stichwörter der Autor:innen

gait analysisknee kinematicsmotion analysisvalidationwearable sensor

Autor:innen und Institutionen

  • Faculty of Health Sciences, Hokkaido University, North 12, West 5, Kita-ku, Sapporo 060-0812, Japan.

ORCID-Profile: Tomoya Ishida, Mina Samukawa

Förderung

  • Nitto (Japan)
  • Nitto Denko Corporation, Tokyo, Japan

Zitieren

APA

Ishida, T., & Samukawa, M. (2024). The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors. Sensors. https://doi.org/10.3390/s24072092

Vancouver

Ishida T, Samukawa M. The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors. Sensors. 2024. doi:10.3390/s24072092

BibTeX
@article{Ishida2024,
  author = {Ishida, Tomoya and Samukawa, Mina},
  title = {The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s24072092}
}
Kostenlos starten →