The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors
Sensors · 2024
DOI: 10.3390/s24072092 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Tomoya Ishida und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Health Sciences, Hokkaido University, North 12, West 5, Kita-ku, Sapporo 060-0812, Japan. Hauptthemen laut MEDLINE: Optical Devices, Wearable Electronic Devices. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nitto (Japan), Nitto Denko Corporation, Tokyo, Japan.
Abstract (Originalsprache)
Frontal and axial knee motion can affect the accuracy of the knee extension/flexion motion measurement using a wearable goniometer. The purpose of this study was to test the hypothesis that calibrating the goniometer on an individual’s body would reduce errors in knee flexion angle during gait, compared to bench calibration. Ten young adults (23.2 ± 1.3 years) were enrolled. Knee flexion angles during gait were simultaneously assessed using a wearable goniometer sensor and an optical three-dimensional motion analysis system, and the absolute error (AE) between the two methods was calculated. The mean AE across a gait cycle was 2.4° (0.5°) for the on-body calibration, and the AE was acceptable (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Health Sciences, Hokkaido University, North 12, West 5, Kita-ku, Sapporo 060-0812, Japan.
ORCID-Profile: Tomoya Ishida, Mina Samukawa
Förderung
- Nitto (Japan)
- Nitto Denko Corporation, Tokyo, Japan
Zitieren
APA
Ishida, T., & Samukawa, M. (2024). The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors. Sensors. https://doi.org/10.3390/s24072092
Vancouver
Ishida T, Samukawa M. The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors. Sensors. 2024. doi:10.3390/s24072092
BibTeX
@article{Ishida2024,
author = {Ishida, Tomoya and Samukawa, Mina},
title = {The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24072092}
}
EUVIMED