A Small Highly Sensitive Glucose Sensor Based on a Glucose Oxidase-Modified U-Shaped Microfiber
Sensors · 2024
DOI: 10.3390/s24020684 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Tingkuo Chen und 3 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Precision Electronic Manufacturing Technology and Equipment, School of Electromechanical Engineering, Guangdong University of…. Hauptthemen laut MEDLINE: Blood Glucose, Biosensing Techniques. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Hongyan Xia, Kang Xie, Science and Technology Planning Project of Guangzhou.
Abstract (Originalsprache)
Diabetes patients need to monitor blood glucose all year round. In this article, a novel scheme is proposed for blood glucose detection. The proposed sensor is based on a U-shaped microfiber prepared using hydrogen-oxygen flame-heating technology, and then 3-aminopropyltriethoxysilane (APTES) and glucose oxidase (GOD) are successively coated on the surface of the U-shaped microfiber via a coating technique. The glucose reacts with the GOD of the sensor surface to produce gluconic acid, which changes the effective refractive index and then shifts the interference wavelength. The structure and morphology of the sensor were characterized via scanning electron microscope (SEM) and confocal laser microscopy (CLM). The experimental results show that the sensitivity of the sensor is as high as 5.73 nm/(mg/mL). Compared with the glucose sensor composed of the same material, the sensitivity of the sensor increased by 329%. The proposed sensor has a broad application prospect in blood glucose detection of diabetic patients due to the advantages of miniaturization, high sensitivity, and good stability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- State Key Laboratory of Precision Electronic Manufacturing Technology and Equipment, School of Electromechanical Engineering, Guangdong University of Technology, Guangzhou 510006, China.
- Department of Applied Sciences, Northumbria University, Newcastle upon Tyne NE1 8ST, UK.
Förderung
- Hongyan Xia
- Kang Xie
- Science and Technology Planning Project of Guangzhou
- Leading Talents of Guangdong Province Program
Zitieren
APA
Chen, T., Jiang, H., Xie, K., & Xia, H. (2024). A Small Highly Sensitive Glucose Sensor Based on a Glucose Oxidase-Modified U-Shaped Microfiber. Sensors. https://doi.org/10.3390/s24020684
Vancouver
Chen T, Jiang H, Xie K, Xia H. A Small Highly Sensitive Glucose Sensor Based on a Glucose Oxidase-Modified U-Shaped Microfiber. Sensors. 2024. doi:10.3390/s24020684
BibTeX
@article{Chen2024,
author = {Chen, Tingkuo and Jiang, Haiming and Xie, Kang and Xia, Hongyan},
title = {A Small Highly Sensitive Glucose Sensor Based on a Glucose Oxidase-Modified U-Shaped Microfiber},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24020684}
}
EUVIMED