Sensor-to-Bone Calibration with the Fusion of IMU and Bi-Plane X-rays

Xavier Gasparutto iD, Kevin Rose-Dulcina iD, Gautier Grouvel iD, Peter DiGiovanni iD, Lena Carcreff iD, Didier Hannouche iD, Stéphane Armand iD

Sensors · 10.01.2024 · 24 (2) · 419

DOI: 10.3390/s24020419 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 10.01.2024. Verfasst von Xavier Gasparutto und 6 weiteren Autor:innen. Beteiligte Einrichtung: Kinesiology Laboratory, Geneva University Hospitals and University of Geneva, 1205 Geneva, Switzerland. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: “Fondation pour la recherche osteo-articulaire” of Geneva, Division of Orthopaedic Surgery and Musculoskeletal Trauma Care of the Geneva University Hospitals.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Inertial measurement units (IMUs) need sensor-to-segment calibration to measure human kinematics. Multiple methods exist, but, when assessing populations with locomotor function pathologies, multiple limitations arise, including holding postures (limited by joint pain and stiffness), performing specific tasks (limited by lack of selectivity) or hypothesis on limb alignment (limited by bone deformity and joint stiffness). We propose a sensor-to-bone calibration based on bi-plane X-rays and a specifically designed fusion box to measure IMU orientation with respect to underlying bones. Eight patients undergoing total hip arthroplasty with bi-plane X-rays in their clinical pathway participated in the study. Patients underwent bi-plane X-rays with fusion box and skin markers followed by a gait analysis with IMUs and a marker-based method. The validity of the pelvis, thigh and hip kinematics measured with a conventional sensor-to-segment calibration and with the sensor-to-bone calibration were compared. Results showed (1) the feasibility of the fusion of bi-plane X-rays and IMUs in measuring the orientation of anatomical axes, and (2) higher validity of the sensor-to-bone calibration for the pelvic tilt and similar validity for other degrees of freedom. The main strength of this novel calibration is to remove conventional hypotheses on joint and segment orientations that are frequently violated in pathological populations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipHumansX-RaysCalibrationRadiographyExtremities

Stichwörter der Autor:innen

IMUbi-plane X-raysfusiongait analysismedical imagingsensor-to-segment calibration

Autor:innen und Institutionen

  • Kinesiology Laboratory, Geneva University Hospitals and University of Geneva, 1205 Geneva, Switzerland.
  • Division of Orthopaedic Surgery and Musculoskeletal Trauma Care, Surgery Department, Geneva University Hospitals and University of Geneva, 1205 Geneva, Switzerland.

ORCID-Profile: Xavier Gasparutto, Kevin Rose-Dulcina, Gautier Grouvel, Peter DiGiovanni, Lena Carcreff, Didier Hannouche, Stéphane Armand

Förderung

  • “Fondation pour la recherche osteo-articulaire” of Geneva
  • Division of Orthopaedic Surgery and Musculoskeletal Trauma Care of the Geneva University Hospitals

Zitieren

APA

Gasparutto, X., Rose-Dulcina, K., Grouvel, G., DiGiovanni, P., Carcreff, L., Hannouche, D., & Armand, S. (2024). Sensor-to-Bone Calibration with the Fusion of IMU and Bi-Plane X-rays. Sensors, 24(2), 419. https://doi.org/10.3390/s24020419

Vancouver

Gasparutto X, Rose-Dulcina K, Grouvel G, DiGiovanni P, Carcreff L, Hannouche D, et al. Sensor-to-Bone Calibration with the Fusion of IMU and Bi-Plane X-rays. Sensors. 2024;24(2):419. doi:10.3390/s24020419

BibTeX
@article{Gasparutto2024,
  author = {Gasparutto, Xavier and Rose-Dulcina, Kevin and Grouvel, Gautier and DiGiovanni, Peter and Carcreff, Lena and Hannouche, Didier and Armand, Stéphane},
  title = {Sensor-to-Bone Calibration with the Fusion of IMU and Bi-Plane X-rays},
  journal = {Sensors},
  year = {2024},
  volume = {24},
  number = {2},
  pages = {419},
  doi = {10.3390/s24020419}
}
Kostenlos starten →