EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

The Use of a Droplet Collar Accessory Attached to a Portable near Infrared Instrument to Identify Methanol Contamination in Whisky

Adam Kolobaric iD, Rebecca Orrell-Trigg iD, Seth Orloff, Vanessa Fraser, James Chapman, Daniel Cozzolino iD

Sensors · 04.11.2023 · 23 (21) · 8969

DOI: 10.3390/s23218969 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 04.11.2023. Verfasst von Adam Kolobaric und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Science, RMIT University, Melbourne 3000, Australia. Hauptthemen laut MEDLINE: Spectroscopy, Near-Infrared, Methanol. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this study was to evaluate the ability of a droplet collar accessory attached to a portable near-infrared (NIR) instrument to characterize the artificial contamination of methanol in commercial whisky samples. Unadulterated samples (n = 12) were purchased from local bottle shops where adulterated samples were created by adding methanol (99% pure methanol) at six levels (0.5%, 1%, 2%, 3%, 4% and 5% v/v) to the commercial whisky samples (controls). Samples were analyzed using a drop collar accessory attached to a MicroNIR Onsite instrument (900–1650 nm). Partial least squares (PLS) cross-validation statistics obtained for the prediction of all levels of methanol (from 0 to 5%) addition were considered adequate when the whole adulteration range was used, coefficient of determination in cross-validation (R2cv: 0.95) and standard error in cross of validation (SECV: 0.35% v/v). The cross-validation statistics were R2cv: 0.97, SECV: 0.28% v/v after the 0.5% and 1% v/v methanol addition was removed. These results showed the ability of using a new sample presentation attachment to a portable NIR instrument to analyze the adulteration of whisky with methanol. However, the low levels of methanol adulteration (0.5 and 1%) were not well predicted using the NIR method evaluated.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Spectroscopy, Near-InfraredMethanolFood ContaminationLeast-Squares Analysis

Stichwörter der Autor:innen

adulterationalcoholinfraredmethanolwhisky

Substanzen

Methanol

Autor:innen und Institutionen

  • School of Science, RMIT University, Melbourne 3000, Australia.
  • Faculty of Science, University of Queensland, Brisbane 4072, Australia.
  • Centre for Nutrition and Food Sciences, Queensland Alliance for Agriculture and Food Innovation(QAAFI), University of Queensland, Brisbane 4072, Australia.

ORCID-Profile: Adam Kolobaric, Rebecca Orrell-Trigg, Daniel Cozzolino

Zitieren

APA

Kolobaric, A., Orrell-Trigg, R., Orloff, S., Fraser, V., Chapman, J., & Cozzolino, D. (2023). The Use of a Droplet Collar Accessory Attached to a Portable near Infrared Instrument to Identify Methanol Contamination in Whisky. Sensors, 23(21), 8969. https://doi.org/10.3390/s23218969

Vancouver

Kolobaric A, Orrell-Trigg R, Orloff S, Fraser V, Chapman J, Cozzolino D. The Use of a Droplet Collar Accessory Attached to a Portable near Infrared Instrument to Identify Methanol Contamination in Whisky. Sensors. 2023;23(21):8969. doi:10.3390/s23218969

BibTeX
@article{Kolobaric2023,
  author = {Kolobaric, Adam and Orrell-Trigg, Rebecca and Orloff, Seth and Fraser, Vanessa and Chapman, James and Cozzolino, Daniel},
  title = {The Use of a Droplet Collar Accessory Attached to a Portable near Infrared Instrument to Identify Methanol Contamination in Whisky},
  journal = {Sensors},
  year = {2023},
  volume = {23},
  number = {21},
  pages = {8969},
  doi = {10.3390/s23218969}
}
Kostenlos starten →