EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor

Zheyuan Pei iD, Maxim Balitskiy, Ryan Thalman iD, Kerry E Kelly

Sensors · 26.08.2023 · 23 (17) · 7444

DOI: 10.3390/s23177444 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 26.08.2023. Verfasst von Zheyuan Pei und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemical Engineering, University of Utah, Salt Lake City, UT 84112, USA. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 61 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: US National Science Foundation, National Science Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Formaldehyde is a known human carcinogen and an important indoor and outdoor air pollutant. However, current strategies for formaldehyde measurement, such as chromatographic and optical techniques, are expensive and labor intensive. Low-cost gas sensors have been emerging to provide effective measurement of air pollutants. In this study, we evaluated eight low-cost electrochemical formaldehyde sensors (SFA30, Sensirion®, Staefa, Switzerland) in the laboratory with a broadband cavity-enhanced absorption spectroscopy as the reference instrument. As a group, the sensors exhibited good linearity of response (R2 > 0.95), low limit of detection (11.3 ± 2.07 ppb), good accuracy (3.96 ± 0.33 ppb and 6.2 ± 0.3% N), acceptable repeatability (3.46% averaged coefficient of variation), reasonably fast response (131–439 s) and moderate inter-sensor variability (0.551 intraclass correlation coefficient) over the formaldehyde concentration range of 0–76 ppb. We also systematically investigated the effects of temperature and relative humidity on sensor response, and the results showed that formaldehyde concentration was the most important contributor to sensor response, followed by temperature, and relative humidity. The results suggest the feasibility of using this low-cost electrochemical sensor to measure formaldehyde concentrations at relevant concentration ranges in indoor and outdoor environments.

Schlagwörter

Stichwörter der Autor:innen

broadband cavity-enhanced absorption spectroscopyformaldehydelow-cost electrochemical sensorsensor evaluation

Autor:innen und Institutionen

  • Department of Chemical Engineering, University of Utah, Salt Lake City, UT 84112, USA.
  • Department of Chemistry, Snow College, Ephraim, UT 84627, USA.

ORCID-Profile: Zheyuan Pei, Ryan Thalman

Förderung

  • US National Science Foundation
  • National Science Foundation

Zitieren

APA

Pei, Z., Balitskiy, M., Thalman, R., & Kelly, K. E. (2023). Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor. Sensors, 23(17), 7444. https://doi.org/10.3390/s23177444

Vancouver

Pei Z, Balitskiy M, Thalman R, Kelly KE. Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor. Sensors. 2023;23(17):7444. doi:10.3390/s23177444

BibTeX
@article{Pei2023,
  author = {Pei, Zheyuan and Balitskiy, Maxim and Thalman, Ryan and Kelly, Kerry E},
  title = {Laboratory Performance Evaluation of a Low-Cost Electrochemical Formaldehyde Sensor},
  journal = {Sensors},
  year = {2023},
  volume = {23},
  number = {17},
  pages = {7444},
  doi = {10.3390/s23177444}
}