Comparative Evaluations on Real-Time Monitoring of Temperature Sensors during Endoscopic Laser Application
Sensors · 30.06.2023 · 23 (13) · 6069
DOI: 10.3390/s23136069 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 30.06.2023. Verfasst von Minh Duc Ta und 4 weiteren Autor:innen. Beteiligte Einrichtung: Industry 4.0 Convergence Bionics Engineering, Marine-Integrated Biomedical Technology Center, Pukyong National University, Busan 48513, Republic of K…. Hauptthemen laut MEDLINE: Hyperthermia, Induced. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea, Korea Medical Device Development Fund.
Abstract (Originalsprache)
Temperature sensors, such as Fiber Bragg Grating (FBG) and thermocouple (TC), have been widely used for monitoring the interstitial tissue temperature during laser irradiation. The aim of the current study was to compare the performance of both FBG and TC in real-time temperature monitoring during endoscopic and circumferential laser treatment on tubular tissue structure. A 600-µm core-diameter diffusing applicator was employed to deliver 980-nm laser light (30 W for 90 s) circumferentially for quantitative evaluation. The tip of the TC was covered with a white tube (W-TC) in order to prevent direct light absorption and to minimize temperature overestimation. The temperature measurements in air demonstrated that the measurement difference in the temperature elevations was around 3.5 °C between FBG and W-TC. Ex vivo porcine liver tests confirmed that the measurement difference became lower (less than 1 °C). Ex vivo porcine esophageal tissue using a balloon-integrated catheter exhibited that both FBG and W-TC consistently showed a comparable trend of temperature measurements during laser irradiation (~2 °C). The current study demonstrated that the white tube-covered TC could be a feasible sensor to monitor interstitial tissue temperature with minimal overestimation during endoscopic laser irradiation. Further in vivo studies on gastroesophageal reflux disease will investigate the performance of the W-TC to monitor the temperature of the esophageal mucosa surface in real-time mode to warrant the safety of endoscopic laser treatment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Industry 4.0 Convergence Bionics Engineering, Marine-Integrated Biomedical Technology Center, Pukyong National University, Busan 48513, Republic of Korea.
- TeCure, Inc., Busan 48548, Republic of Korea.
- Division of Smart Healthcare and Digital Healthcare Research Center, College of Information Technology and Convergence, Pukyong National University, Busan 48513, Republic of Korea.
- Marine-Integrated Biomedical Technology Center, The National Key Research Institutes in Universities, Pukyong National University, Busan 48513, Republic of Korea.
ORCID-Profile: Hyun Wook Kang
Förderung
- National Research Foundation of Korea
- Korea Medical Device Development Fund
Zitieren
APA
Ta, M. D., Truong, V. G., Lim, S., Lee, B. I., & Kang, H. W. (2023). Comparative Evaluations on Real-Time Monitoring of Temperature Sensors during Endoscopic Laser Application. Sensors, 23(13), 6069. https://doi.org/10.3390/s23136069
Vancouver
Ta MD, Truong VG, Lim S, Lee BI, Kang HW. Comparative Evaluations on Real-Time Monitoring of Temperature Sensors during Endoscopic Laser Application. Sensors. 2023;23(13):6069. doi:10.3390/s23136069
BibTeX
@article{Ta2023,
author = {Ta, Minh Duc and Truong, Van Gia and Lim, Seonghee and Lee, Byeong-Il and Kang, Hyun Wook},
title = {Comparative Evaluations on Real-Time Monitoring of Temperature Sensors during Endoscopic Laser Application},
journal = {Sensors},
year = {2023},
volume = {23},
number = {13},
pages = {6069},
doi = {10.3390/s23136069}
}