EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Comparison of quartz and sapphire optical chambers for infrared laser sealing of vascular tissues using a reciprocating, side‐firing optical fiber: Simulations and experiments

Woheeb M Saeed, Patrick J O'Brien, Jude Yoshino, Aidan R Restelli, Alexandria J Traynham, Nathaniel M Fried iD

Lasers in Surgery and Medicine · 2023

DOI: 10.1002/lsm.23740 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Lasers in Surgery and Medicine, 2023. Verfasst von Woheeb M Saeed und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Optical Science, University of North Carolina at Charlotte, Charlotte, North Carolina, USA. Hauptthemen laut MEDLINE: Optical Fibers, Laparoscopy. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Infrared (IR) lasers are being tested as an alternative to radiofrequency (RF) and ultrasonic (US) surgical devices for hemostatic sealing of vascular tissues. In previous studies, a side‐firing optical fiber with elliptical IR beam output was reciprocated, producing a linear IR laser beam pattern for uniform sealing of blood vessels. Technical challenges include limited field‐of‐view of vessel position within the metallic device jaws, and matching fiber scan length to variable vessel sizes. A transparent jaw may improve visibility and enable custom treatment.

Methods

Quartz and sapphire square optical chambers (2.7 × 2.7 × 25 [mm 3 ] outer dimensions) were tested, capable of fitting into a 5‐mm‐OD laparoscopic device. A 1470 nm laser was used for optical transmission studies. Razor blade scans and an IR beam profiler acquired fiber (550‐µm‐core/0.22NA) output beam profiles. Thermocouples recorded peak temperatures and cooling times on internal and external chamber surfaces. Optical fibers with angle polished distal tips delivered 94% of light at a 90° angle. Porcine renal arteries with diameters of 3.4 ± 0.7 mm ( n = 13) for quartz and 3.2 ± 0.7 mm ( n = 14) for sapphire chambers ( p > 0.05), were sealed using 30 W for 5 s.

Results

Reflection losses at material/air interfaces were 3.3% and 7.4% for quartz and sapphire. Peak temperatures on the external chamber surface averaged 74 ± 8°C and 73 ± 10°C ( p > 0.05). Times to cool down to 37°C measured 13 ± 4 s and 27 ± 7 s ( p < 0.05). Vessel burst pressures (BP) averaged 883 ± 393 mmHg and 412 ± 330 mmHg ( p < 0.05). For quartz, 13/13 (100%) vessels were sealed (BP > 360 mmHg), versus 9/14 (64%) for sapphire. Computer simulations for the quartz chamber yielded peak temperatures (78°C) and cooling times (16 s) similar to experiments.

Conclusions

Quartz is an inexpensive material for use in a laparoscopic device jaw, providing more consistent vessel seals and faster cooling times than sapphire and current RF and US devices.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Optical FibersLaparoscopySwineAnimalsQuartzAluminum OxideLasers

Stichwörter der Autor:innen

blood vesselscoagulationfusioninfrared laserlaparoscopicsealing

Substanzen

QuartzAluminum Oxide

Autor:innen und Institutionen

  • Department of Physics and Optical Science, University of North Carolina at Charlotte, Charlotte, North Carolina, USA.
  • Department of Mechanical Engineering, University of North Carolina at Charlotte, Charlotte, North Carolina, USA.

ORCID-Profile: Nathaniel M Fried

Förderung

  • NIBIB NIH HHS

Zitieren

APA

Saeed, W. M., O'Brien, P. J., Yoshino, J., Restelli, A. R., Traynham, A. J., & Fried, N. M. (2023). Comparison of quartz and sapphire optical chambers for infrared laser sealing of vascular tissues using a reciprocating, side‐firing optical fiber: Simulations and experiments. Lasers in Surgery and Medicine. https://doi.org/10.1002/lsm.23740

Vancouver

Saeed WM, O'Brien PJ, Yoshino J, Restelli AR, Traynham AJ, Fried NM. Comparison of quartz and sapphire optical chambers for infrared laser sealing of vascular tissues using a reciprocating, side‐firing optical fiber: Simulations and experiments. Lasers in Surgery and Medicine. 2023. doi:10.1002/lsm.23740

BibTeX
@article{Saeed2023,
  author = {Saeed, Woheeb M and O'Brien, Patrick J and Yoshino, Jude and Restelli, Aidan R and Traynham, Alexandria J and Fried, Nathaniel M},
  title = {Comparison of quartz and sapphire optical chambers for infrared laser sealing of vascular tissues using a reciprocating, side‐firing optical fiber: Simulations and experiments},
  journal = {Lasers in Surgery and Medicine},
  year = {2023},
  doi = {10.1002/lsm.23740}
}
Kostenlos starten →