EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Functional Performance and Patient-Reported Goal Attainment with a Novel Microprocessor-Controlled Knee in Active Transfemoral Amputees: A Multicenter Randomized Crossover Trial

Victoria Spartacus, Marta Da Costa, Martin Baudry, Lisa Gaffino, Sandrine Rey, Clément Duraffourg, Isabelle Loiret, Laurine Calistri

Prosthesis · 04.08.2026 · 8 (8) · 82

DOI: 10.3390/prosthesis8080082 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Prosthesis, online erschienen am 04.08.2026. Verfasst von Victoria Spartacus und 7 weiteren Autor:innen. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: PROTEOR.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Lower-limb amputation significantly impacts autonomy and daily functioning. Microprocessor-controlled prosthetic knees (MPKs) improve mobility, safety, and quality of life compared with non-microprocessor prosthetic knees (nMPKs). Despite general benefits, differences among MPK models remain unclear. The goal of this investigation was to evaluate functional mobility and patient-reported outcomes in active transfemoral amputees using a new MPK (PROTEOR QUATTRO) compared with existing MPKs.

Methods

In this multicenter cross-over clinical trial, active transfemoral amputees were assessed using patient-specific functional scales (PSFS), standardized locomotor capacity tests (6MWT, 10mWT, HAI, SAI) and satisfaction and quality-of-life questionnaires (PEQ-A, SF-36, ESAT). Biomechanical data collection was performed during walking test, to examine gait symmetry indices.

Results

The QUATTRO MPK was associated with significantly higher PSFS scores, reflecting higher self-rated ability in performing chosen daily life activities compared with patient’s current prosthesis. Standardized locomotor tests revealed comparable performance between devices. Biomechanical analyses showed similar symmetry indices for QUATTRO and other MPKs. Patient-reported outcomes showed an improvement in physical health with QUATTRO, while overall satisfaction remained similar.

Conclusions

QUATTRO improved patient-perceived ability to perform meaningful daily-life activities while maintaining comparable performance on standardized locomotor tests. These results highlight the importance of patient-centered assessments in identifying functional benefits that may guide individualized prosthetic prescriptions for active transfemoral amputees.

Förderung

  • PROTEOR

Zitieren

APA

Spartacus, V., Costa, M. D., Baudry, M., Gaffino, L., Rey, S., Duraffourg, C., Loiret, I., & Calistri, L. (2026). Functional Performance and Patient-Reported Goal Attainment with a Novel Microprocessor-Controlled Knee in Active Transfemoral Amputees: A Multicenter Randomized Crossover Trial. Prosthesis, 8(8), 82. https://doi.org/10.3390/prosthesis8080082

Vancouver

Spartacus V, Costa MD, Baudry M, Gaffino L, Rey S, Duraffourg C, et al. Functional Performance and Patient-Reported Goal Attainment with a Novel Microprocessor-Controlled Knee in Active Transfemoral Amputees: A Multicenter Randomized Crossover Trial. Prosthesis. 2026;8(8):82. doi:10.3390/prosthesis8080082

BibTeX
@article{Spartacus2026,
  author = {Spartacus, Victoria and Costa, Marta Da and Baudry, Martin and Gaffino, Lisa and Rey, Sandrine and Duraffourg, Clément and Loiret, Isabelle and Calistri, Laurine},
  title = {Functional Performance and Patient-Reported Goal Attainment with a Novel Microprocessor-Controlled Knee in Active Transfemoral Amputees: A Multicenter Randomized Crossover Trial},
  journal = {Prosthesis},
  year = {2026},
  volume = {8},
  number = {8},
  pages = {82},
  doi = {10.3390/prosthesis8080082}
}