Toward personalizing prosthesis prescription: A take‐home study of three microprocessor‐controlled prosthetic knees: A randomized crossover study
PM&R · 2026-03 · 18 (3) · 288-304
DOI: 10.1002/pmrj.70028 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift PM&R, online erschienen am 24.10.2025. Verfasst von Kinsey Herrin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, Georgia, USA. Hauptthemen laut MEDLINE: Knee Prosthesis, Microcomputers, Amputation, Surgical, Amputees. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Congressionally Directed Medical Research Programs.
Abstract (Originalsprache)
Background
Previous studies on microprocessor‐controlled prosthetic knees (MPKs) often investigate benefits of MPKs as a class of knees rather than clinically relevant differences between specific knees, despite their distinct features.
Objectives
To systematically evaluate and report outcomes associated with three commercially available MPKs following a standardized real‐world use period.
Design
Randomized crossover study.
Setting
Research laboratory and community environment.
Participants
Ten patients with transfemoral amputation.
Interventions
Three MPKs were fitted, trained, and worn for a 1‐week period including C‐Leg 4.0 (Ottobock, Duderstadt, Germany), Rheo Knee‐Model RM7 (Össur, Reykjavik, Iceland), and Power Knee‐PKA01 (Össur, Reykjavik, Iceland). Main Outcome Measures Primary outcomes were the 10‐meter walk test (10‐mwt), the 2‐minute walk test (2‐mwt), and the Prosthesis Evaluation Questionnaire (PEQ). Secondary outcomes were stance time asymmetry, physiological cost index, stair and ramp speeds, the narrowing beam walking test, and community ambulation monitoring.
Results
Participants walked 11% faster in Rheo than Power Knee during the 10‐mwt (95% confidence interval [CI]: 0.046–0.184, p = .015). In the 2‐mwt, participants walked 12% faster in C‐Leg (95% CI: 0.034–0.241, p = .003) and 9% faster in Rheo (95% CI: 0.031, 0.163, p = .027) than in Power Knee. On the PEQ, participants reported greater satisfaction with C‐Leg compared to Power Knee ( p = .006). Ramp ascent speed was 8% faster in Rheo than Power Knee (95% CI: 0.026–0.130, p = .024). No significant differences were found for other secondary outcomes. Notably, 10 of 12 outcomes showed individuals performing their best by a defined difference on an MPK different from the cohort's best‐performing MPK.
Conclusions
Participants walked faster in C‐Leg and Rheo than Power Knee and reported greater satisfaction with C‐Leg. Consideration of patient needs and characteristics may allow more individualized MPK prescription and thereby improve rehabilitation outcomes. Database Registration NCT06399471.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, Georgia, USA.
- Institute for Robotics and Intelligent Machines, Georgia Institute of Technology, Atlanta, Georgia, USA.
- School of Biological Sciences, Georgia Institute of Technology, Atlanta, Georgia, USA.
- Extremity Trauma and Amputation Center of Excellence (EACE), Military Performance Lab, Center for the Intrepid, Brooke Army Medical Center, San Antonio, Texas, USA.
- Center for the Intrepid, Department of Rehabilitation Medicine, Brooke Army Medical Center, San Antonio, Texas, USA.
ORCID-Profile: Kinsey Herrin
Förderung
- Congressionally Directed Medical Research Programs
Zitieren
APA
Herrin, K., Kestur, S., Zhou, S., O'Sullivan, G., Snow, T., Childers, W. L., & Young, A. (2026). Toward personalizing prosthesis prescription: A take‐home study of three microprocessor‐controlled prosthetic knees: A randomized crossover study. PM&R, 18(3), 288-304. https://doi.org/10.1002/pmrj.70028
Vancouver
Herrin K, Kestur S, Zhou S, O'Sullivan G, Snow T, Childers WL, et al. Toward personalizing prosthesis prescription: A take‐home study of three microprocessor‐controlled prosthetic knees: A randomized crossover study. PM&R. 2026;18(3):288-304. doi:10.1002/pmrj.70028
BibTeX
@article{Herrin2026,
author = {Herrin, Kinsey and Kestur, Sujay and Zhou, Sixu and O'Sullivan, Gwyn and Snow, Teresa and Childers, Walter Lee and Young, Aaron},
title = {Toward personalizing prosthesis prescription: A take‐home study of three microprocessor‐controlled prosthetic knees: A randomized crossover study},
journal = {PM&R},
year = {2026},
volume = {18},
number = {3},
pages = {288--304},
doi = {10.1002/pmrj.70028}
}
EUVIMED