Comparative Fatigue Analysis of CF-PLA Metamaterial Bone Plates for Orthopaedic Fixation

Ani Daniel iD, Hamed Bakhtiari iD, Barun K Das, Muhammad Aamir iD, Majid Tolouei-Rad iD

Polymers · 08.05.2026 · 18 (10) · 1152

DOI: 10.3390/polym18101152 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Polymers, online erschienen am 08.05.2026. Verfasst von Ani Daniel und 4 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Advanced Materials and Manufacturing (CAMM), School of Engineering, Edith Cowan University, Joondalup, WA 6027, Australia. Die Arbeit nennt 90 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Edith Cowan University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Bone plates are widely used in orthopaedic surgery to stabilise fractured bones and support healing following traumatic injuries or osteotomies. However, conventional metallic bone plates suffer from stress shielding and stiffness mismatch with bone, which can hinder optimal healing. Additive manufacturing enables the incorporation of novel metamaterial architectures into polymer-based implants to enhance mechanical properties. The fatigue behaviour of these implants during the healing period is critical to ensuring their structural integrity and long-term performance. In this study, the compressive fatigue performance of fused deposition modelling (FDM)-printed carbon fibre-reinforced polylactic acid (CF-PLA) bone plates were investigated. Four metamaterial structures—tetrachiral, re-entrant, rotating square, and hexagonal—were evaluated under strain-controlled cyclic loading at 20%, 40%, 60%, and 80% of their respective yield strains. The results showed a strong dependence of fatigue behaviour on lattice geometry. Among the tested configurations, the re-entrant structured bone plate exhibited the best overall fatigue performance, sustaining up to 100,000 cycles at moderate strain levels and showing delayed stiffness degradation under high strain conditions. In contrast, rotating square and hexagonal structures showed early stiffness loss and failure at higher strain levels. These findings highlight the importance of lattice design in fatigue performance, although FDM-induced printing defects significantly influence overall fatigue behaviour.

Schlagwörter

Stichwörter der Autor:innen

3D printingCF-PLACFRPFDMbone platefatigue behaviourmetamaterialspolymers

Autor:innen und Institutionen

  • Centre for Advanced Materials and Manufacturing (CAMM), School of Engineering, Edith Cowan University, Joondalup, WA 6027, Australia.
  • School of Engineering, Edith Cowan University, Joondalup, WA 6027, Australia.

ORCID-Profile: Ani Daniel, Hamed Bakhtiari, Muhammad Aamir, Majid Tolouei-Rad

Förderung

  • Edith Cowan University

Zitieren

APA

Daniel, A., Bakhtiari, H., Das, B. K., Aamir, M., & Tolouei-Rad, M. (2026). Comparative Fatigue Analysis of CF-PLA Metamaterial Bone Plates for Orthopaedic Fixation. Polymers, 18(10), 1152. https://doi.org/10.3390/polym18101152

Vancouver

Daniel A, Bakhtiari H, Das BK, Aamir M, Tolouei-Rad M. Comparative Fatigue Analysis of CF-PLA Metamaterial Bone Plates for Orthopaedic Fixation. Polymers. 2026;18(10):1152. doi:10.3390/polym18101152

BibTeX
@article{Daniel2026,
  author = {Daniel, Ani and Bakhtiari, Hamed and Das, Barun K and Aamir, Muhammad and Tolouei-Rad, Majid},
  title = {Comparative Fatigue Analysis of CF-PLA Metamaterial Bone Plates for Orthopaedic Fixation},
  journal = {Polymers},
  year = {2026},
  volume = {18},
  number = {10},
  pages = {1152},
  doi = {10.3390/polym18101152}
}
Kostenlos starten →