This paper presents a dual-axis micro-displacement measurement system based on double-layer grating interference adopting a quasi-common-mode optical configuration. Utilizing the diffraction and interference effects of the grating pair, the system realizes decoupled high-precision dual-axis displacement measurement by detecting diffracted light intensity induced by out-of-plane and in-plane motions of the movable grating. The light-intensity response originating from grating-pair diffraction is theoretically derived to establish the quantitative relationship between displacement and optical intensity. An experimental setup is constructed to record intensity variation curves for out-of-plane and in-plane motions. Measurement accuracy along the two axes is systematically analyzed over a 1 mm working range. Experimental results yield a full-scale accuracy of 0.0039% and a resolution of 0.38 nm for out-of-plane displacement, while the in-plane displacement measurement achieves a full-scale accuracy of 0.08% with 19 nm resolution. Benefiting from its compact architecture, nanoscale resolution, and superior measurement precision, the proposed grating-pair scheme opens a new avenue for multi-dimensional micro-displacement sensing.
Förderung
- Fundamental Research Program of Shanxi Province
Zitieren
APA
Li, K., Jin, L., Gong, Y., Zhang, C., Xie, K., & Li, M. (2026). Dual-Axis Micro-Displacement Sensing Based on a Grating Pair with a Quasi-Common-Mode Optical Configuration. Micromachines, 17(11), 1218. https://doi.org/10.3390/mi17111218
Vancouver
Li K, Jin L, Gong Y, Zhang C, Xie K, Li M. Dual-Axis Micro-Displacement Sensing Based on a Grating Pair with a Quasi-Common-Mode Optical Configuration. Micromachines. 2026;17(11):1218. doi:10.3390/mi17111218
BibTeX
@article{Li2026,
author = {Li, Kexin and Jin, Li and Gong, Yancheng and Zhang, Chao and Xie, Kunyang and Li, Mengwei},
title = {Dual-Axis Micro-Displacement Sensing Based on a Grating Pair with a Quasi-Common-Mode Optical Configuration},
journal = {Micromachines},
year = {2026},
volume = {17},
number = {11},
pages = {1218},
doi = {10.3390/mi17111218}
}
EUVIMED