EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Finite Element Analysis and Computational Framework for Optimizing Laser Surface Modified Ti-6Al-4V Femoral Components in Total Knee Replacement

Iman Shakir Tawfeeq, Hussam Lefta Alwan, Taha A Elwi iD

Micromachines · 18.06.2026 · 17 (6) · 740

DOI: 10.3390/mi17060740 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Micromachines, online erschienen am 18.06.2026. Verfasst von Iman Shakir Tawfeeq und 2 weiteren Autor:innen. Beteiligte Einrichtung: College of Electromechanically Engineering, University of Technology, Baghdad 10066, Iraq. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Titanium alloys such as Ti-6Al-4V are widely used in orthopedic implants due to their strength and biocompatibility. Laser Surface Remelting (LSR) offers a promising approach to modify surface properties without altering bulk characteristics. This study investigates the effects of varying melt pool depths (MPDs) from 0 μm to 30 μm in 10 μm steps on the mechanical behavior of Ti-6Al-4V femoral components in Total Knee Replacement (TKR) using a comprehensive computational approach combining Finite Element Analysis (FEA) and computational algorithm-based automated evaluation. A three-dimensional FEA model was developed and tested under four physiological loading conditions: compression, axial distraction, medial bending, and lateral flexion at 1300 N. Results show that increasing MPD from 0 μm to 30 μm increases the maximum von Mises stress by 4.2% under compression but reduces displacement by up to 51.7% under distraction. An MPD of 20 μm reduces displacement by 48% while increasing stress by only 2.7%, representing an optimal balance. The computational algorithm framework identifies 15–25 μm as the optimal range for balancing surface enhancement with mechanical integrity. Experimental validation shows good agreement between simulated and measured results, confirming the reliability of the proposed framework for optimizing surface modification parameters in orthopedic implants.

Schlagwörter

Stichwörter der Autor:innen

FEAGUILSRMPDTKRTi-6Al-4V

Autor:innen und Institutionen

  • College of Electromechanically Engineering, University of Technology, Baghdad 10066, Iraq.
  • College of Production Engineering and Metallurgy, University of Technology, Baghdad 10066, Iraq.
  • Department of Automation and Artificial Intelligence Engineering, College of Information Engineering, Al-Nahrain University, Baghdad 64074, Iraq.

ORCID-Profile: Taha A Elwi

Zitieren

APA

Tawfeeq, I. S., Alwan, H. L., & Elwi, T. A. (2026). Finite Element Analysis and Computational Framework for Optimizing Laser Surface Modified Ti-6Al-4V Femoral Components in Total Knee Replacement. Micromachines, 17(6), 740. https://doi.org/10.3390/mi17060740

Vancouver

Tawfeeq IS, Alwan HL, Elwi TA. Finite Element Analysis and Computational Framework for Optimizing Laser Surface Modified Ti-6Al-4V Femoral Components in Total Knee Replacement. Micromachines. 2026;17(6):740. doi:10.3390/mi17060740

BibTeX
@article{Tawfeeq2026,
  author = {Tawfeeq, Iman Shakir and Alwan, Hussam Lefta and Elwi, Taha A},
  title = {Finite Element Analysis and Computational Framework for Optimizing Laser Surface Modified Ti-6Al-4V Femoral Components in Total Knee Replacement},
  journal = {Micromachines},
  year = {2026},
  volume = {17},
  number = {6},
  pages = {740},
  doi = {10.3390/mi17060740}
}