Mechanics of Long-Shank 5 mm Neural Probe Insertion into the Rat Brain: Effects of Geometry and Vibration-Assisted Insertion
Micromachines · 31.05.2026 · 17 (6) · 684
DOI: 10.3390/mi17060684 ↗Kurz gefasst
Fachartikel in der Zeitschrift Micromachines, online erschienen am 31.05.2026. Verfasst von Mahasty Khajehzadeh und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bioengineering, The University of Texas at Dallas, Richardson, TX 75080, USA. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, NINDS NIH HHS, National Institute of Neurological Disorders.
Abstract (Originalsprache)
Insertion of microelectrode arrays (MEAs) into brain tissue remains a mechanical challenge, especially for long, thin probes designed to access deep structures. This study investigates the mechanical properties of 5 mm long amorphous silicon carbide (a-SiC) probes with different geometries and the effect of vibration-assisted insertion on penetration into rat brain.
Methods
Two planar a-SiC probe designs were fabricated with identical lengths and thicknesses but differing width geometries: one with a uniform width (175 µm) and the other with a tapered shape (tapering from 175 to 75 µm). Critical buckling forces (FCs) were estimated by finite element modeling (FEM) and validated experimentally. Insertion mechanics were assessed in a brain mimic of 1.2% agarose gel at varying insertion speeds (20–1000 µm/s) and in vivo by implantation in rat cortex. Insertion metrics included penetration force (FP), cortical dimpling depth (Dd), maximum insertion force (Fmax), and success rate of insertion, all evaluated with and without vibrational assistance.
Results
The tapered design exhibited lower penetration force and higher insertion success compared to the uniform-width probe, despite having a lower critical buckling force. An optimal insertion rate of 100 µm/s was identified, balancing insertion time with low Fmax and high insertion success across designs. Higher FP and Dd with a lower success rate were observed for uniform probes compared with tapered probes in rat brain. Vibration-assisted insertion was then investigated with tapered probes. Applying vibration significantly reduced FP, whereas Dd and Fmax remained unchanged. Notably, in 40% of actuated insertions in rat, no detectable FP peak was observed, suggesting unimpeded pial penetration.
Conclusions
A tapered probe geometry and vibration-assisted insertion can reduce Fmax and FP while enhancing the insertion success rate for probe penetration in rat brain. These strategies are generally applicable to long-shank MEA insertions in brain and may inform design and insertion strategies.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Bioengineering, The University of Texas at Dallas, Richardson, TX 75080, USA.
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH 44106, USA.
- Department of Electrical and Computer Engineering, University of Utah, Salt Lake City, UT 84112, USA.
- Office of Research and Innovation, The University of Texas at Dallas, Richardson, TX 75080, USA.
- Division of Child Neurology, Department of Pediatrics, Neurology, Neuroscience and Philosophy, Weill Cornell Medicine, New York, NY 10065, USA.
ORCID-Profile: Mahasty Khajehzadeh, Christopher K Nguyen, Mrigank Maharana, Alexandra Joshi-Imre, Juan M Pascual, Stuart F Cogan
Förderung
- NIH HHS
- NINDS NIH HHS
- National Institute of Neurological Disorders
- Stroke of the National Institutes of Health
Zitieren
APA
Khajehzadeh, M., Nguyen, C. K., Maharana, M., Peddapuram, S., Joshi-Imre, A., Pascual, J. M., & Cogan, S. F. (2026). Mechanics of Long-Shank 5 mm Neural Probe Insertion into the Rat Brain: Effects of Geometry and Vibration-Assisted Insertion. Micromachines, 17(6), 684. https://doi.org/10.3390/mi17060684
Vancouver
Khajehzadeh M, Nguyen CK, Maharana M, Peddapuram S, Joshi-Imre A, Pascual JM, et al. Mechanics of Long-Shank 5 mm Neural Probe Insertion into the Rat Brain: Effects of Geometry and Vibration-Assisted Insertion. Micromachines. 2026;17(6):684. doi:10.3390/mi17060684
BibTeX
@article{Khajehzadeh2026,
author = {Khajehzadeh, Mahasty and Nguyen, Christopher K and Maharana, Mrigank and Peddapuram, Shriya and Joshi-Imre, Alexandra and Pascual, Juan M and Cogan, Stuart F},
title = {Mechanics of Long-Shank 5 mm Neural Probe Insertion into the Rat Brain: Effects of Geometry and Vibration-Assisted Insertion},
journal = {Micromachines},
year = {2026},
volume = {17},
number = {6},
pages = {684},
doi = {10.3390/mi17060684}
}