Design and Control of Upper Limb Rehabilitation Training Robot Based on a Magnetorheological Joint Damper

Jintao Zhu, Hongsheng Hu, Wei Zhao, Jiabin Yang, Qing Ouyang

Micromachines · 22.02.2024 · 15 (3) · 301

DOI: 10.3390/mi15030301 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Micromachines, online erschienen am 22.02.2024. Verfasst von Jintao Zhu und 4 weiteren Autor:innen. Beteiligte Einrichtung: College of Mechanical Engineering, Zhejiang University of Technology, Hangzhou 310023, China. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Jiaxing Municipal Science and Technology Project, Zhejiang Provincial Natural Science Foundation of China, Natural Science Foundation of China (NSFC) grant funded by the Chinese Government.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In recent years, rehabilitation robots have been developed and used in rehabilitation training for patients with hemiplegia. In this paper, a rehabilitation training robot with variable damping is designed to train patients with hemiplegia to recover upper limb function. Firstly, a magnetorheological joint damper (MR joint damper) is designed for the rehabilitation training robot, and its structural design and dynamic model are tested theoretically and experimentally. Secondly, the rehabilitation robot is simplified into a spring-damping system, and the rehabilitation training controller for human movement is designed. The rehabilitation robot dynamically adjusts the excitation current according to the feedback speed and human–machine interaction torque, so that the rehabilitation robot always outputs a stable torque. The magnetorheological joint damper acts as a clutch to transmit torque safely and stably to the robot joint. Finally, the upper limb rehabilitation device is tested. The expected torque is set to 20 N, and the average value of the output expected torque during operation is 20.02 N, and the standard deviation is 0.635 N. The output torque has good stability. A fast (0.5 s) response can be achieved in response to a sudden motor speed change, and the average expected output torque is 20.38 N and the standard deviation is 0.645 N, which can still maintain the stability of the output torque.

Schlagwörter

Stichwörter der Autor:innen

magnetorheological joint damperrehabilitation trainingsafety testingtrajectory tracking controlupper limb rehabilitation robot

Autor:innen und Institutionen

  • College of Mechanical Engineering, Zhejiang University of Technology, Hangzhou 310023, China.
  • College of Information Science and Engineering, Jiaxing University, Jiaxing 314001, China.
  • School of Mechanical Engineering, Nanjing University of Science and Technology, Nanjing 210044, China.
  • Taizhou Jiuju Technology Co., Ltd., Taizhou 318000, China.

Förderung

  • Jiaxing Municipal Science and Technology Project
  • Zhejiang Provincial Natural Science Foundation of China
  • Natural Science Foundation of China (NSFC) grant funded by the Chinese Government
  • Qin Shen Scholar Program of Jiaxing University

Zitieren

APA

Zhu, J., Hu, H., Zhao, W., Yang, J., & Ouyang, Q. (2024). Design and Control of Upper Limb Rehabilitation Training Robot Based on a Magnetorheological Joint Damper. Micromachines, 15(3), 301. https://doi.org/10.3390/mi15030301

Vancouver

Zhu J, Hu H, Zhao W, Yang J, Ouyang Q. Design and Control of Upper Limb Rehabilitation Training Robot Based on a Magnetorheological Joint Damper. Micromachines. 2024;15(3):301. doi:10.3390/mi15030301

BibTeX
@article{Zhu2024,
  author = {Zhu, Jintao and Hu, Hongsheng and Zhao, Wei and Yang, Jiabin and Ouyang, Qing},
  title = {Design and Control of Upper Limb Rehabilitation Training Robot Based on a Magnetorheological Joint Damper},
  journal = {Micromachines},
  year = {2024},
  volume = {15},
  number = {3},
  pages = {301},
  doi = {10.3390/mi15030301}
}
Kostenlos starten →