Performance of XL Sizes of Myval Balloon-Expandable Valve in Real-World Patients with Extremely Large Aortic Annuli
Medicina · 20.03.2026 · 62 (3) · 585
DOI: 10.3390/medicina62030585 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Medicina, online erschienen am 20.03.2026. Verfasst von Kasparas Briedis und 16 weiteren Autor:innen. Beteiligte Einrichtung: Heart Centre, Lithuanian University of Health Sciences, 44307 Kaunas, Lithuania. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Aortic Valve Stenosis, Heart Valve Prosthesis, Aortic Valve. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background and Objectives: Transcatheter aortic valve replacement (TAVR) in large aortic annuli poses challenges due to limited valve-size options and increased complication risks. The aim is to evaluate the safety and performance of XL sizes (30.5 mm and 32 mm) of the Myval transcatheter heart valve (THV) for treating patients with severe aortic stenosis and large aortic annuli. Material and Methods: This retrospective observational study included consecutive patients undergoing TAVR with XL sizes of the Myval THV between December 2023 and December 2024 at a single centre. During this period, 146 TAVI procedures were performed, of which 15 patients (10.3%) with large aortic annuli (mean systolic annular area 786.5 ± 48.2 mm2) received XL valves and were included in the present analysis. Patients were followed up at discharge, 3–6 months, and 1 year. Patient evaluation included echocardiography and clinical assessments following the Valve Academic Research Consortium-3 criteria.
Results
All patients were male, with a mean age of 79.1 ± 5.9 years. Technical success was achieved in 100% of cases. At discharge, none of the patients had moderate or greater paravalvular leakage (PVL); 11 patients had no PVL, while 1 had trace and 3 had mild PVL. The mean effective orifice area (EOA) improved from 0.75 ± 0.15 cm2 at baseline to 2.31 ± 0.21 cm2 at discharge (p < 0.0001). At the 12-month follow-up, the mean EOA was 2.4 ± 0.3 cm2, and no moderate or severe PVL or major adverse clinical outcomes were observed. One patient required a permanent pacemaker implantation due to an atrioventricular block.
Conclusions
The XL sizes of Myval THV showed both safety and efficacy in patients with large aortic annuli, demonstrating acceptable hemodynamic performance and low complication rates. However, large-scale studies with longer follow-ups are needed to validate these findings in diverse populations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart Centre, Lithuanian University of Health Sciences, 44307 Kaunas, Lithuania.
- Medical Academy, Lithuanian University of Health Sciences, 44307 Kaunas, Lithuania.
- Boston Scientific, Ballybrit Business Park, H91 Y868 Galway, Ireland.
- University Hospital Limerick, V94 F858 Limerick, Ireland.
- Bon Secours Hospital Limerick, V94 892F Limerick, Ireland.
ORCID-Profile: Kasparas Briedis, Sandra Kmitaitė, Kamilija Briedė, Ali Aldujeli
Zitieren
APA
Briedis, K., Morkūnaitė, K., Zapustas, N., Zarambaitė, E., Krivickas, Ž., Kmitaitė, S., Rimkutė, A., Tvaronavičiūtė, K., Briedė, K., Lukauskaitė, U., Biesevičienė, M., Tsai, T. Y., Aldujeli, A., Plisienė, J., Unikas, R., Žaliūnas, R., & Bardauskienė, L. (2026). Performance of XL Sizes of Myval Balloon-Expandable Valve in Real-World Patients with Extremely Large Aortic Annuli. Medicina, 62(3), 585. https://doi.org/10.3390/medicina62030585
Vancouver
Briedis K, Morkūnaitė K, Zapustas N, Zarambaitė E, Krivickas Ž, Kmitaitė S, et al. Performance of XL Sizes of Myval Balloon-Expandable Valve in Real-World Patients with Extremely Large Aortic Annuli. Medicina. 2026;62(3):585. doi:10.3390/medicina62030585
BibTeX
@article{Briedis2026,
author = {Briedis, Kasparas and Morkūnaitė, Kristina and Zapustas, Norvydas and Zarambaitė, Evelina and Krivickas, Žilvinas and Kmitaitė, Sandra and Rimkutė, Agnė and Tvaronavičiūtė, Klaudija and Briedė, Kamilija and Lukauskaitė, Urtė and Biesevičienė, Monika and Tsai, Tsung-Ying and Aldujeli, Ali and Plisienė, Jurgita and Unikas, Ramūnas and Žaliūnas, Remigijus and Bardauskienė, Lina},
title = {Performance of XL Sizes of Myval Balloon-Expandable Valve in Real-World Patients with Extremely Large Aortic Annuli},
journal = {Medicina},
year = {2026},
volume = {62},
number = {3},
pages = {585},
doi = {10.3390/medicina62030585}
}
EUVIMED