Pitfalls with the MAKO Robotic-Arm-Assisted Total Knee Arthroplasty
Medicina · 02.02.2024 · 60 (2) · 262
DOI: 10.3390/medicina60020262 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medicina, online erschienen am 02.02.2024. Verfasst von Konstantinos Dretakis und 1 weiteren Autor:innen. Beteiligte Einrichtung: 2nd Department of Orthopaedics, "Hygeia" General Hospital of Athens, 15123 Athens, Greece. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures. Die Arbeit wurde bisher 43-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Robotic-arm-assisted total knee arthroplasty (RATKA) with the MAKO system minimizes deviations in implant alignment and yields superior precision in implant position compared to a manual total knee arthroplasty. In this comprehensive commentary, we present and categorize the limitations and pitfalls of the procedure and we also provide recommendations for avoiding each limitation. The main surgeon-related limitations include prolonged operation duration, loose insertion of the checkpoints and pins, wrong registration and mapping, and damage to soft tissues during bone cutting. The system-related issues include the interruptions of the saw-cutting due to vibrations, specifications for the operating room floor and power supply, the high cost of the system, as well as the cost of each operation due to the extra implants, inability to use the system with various prostheses, wireless connection interruptions between the system’s components, and hardware issues with the six joints of the robotic device. In order to circumvent the potential challenges in this surgical procedure, it is essential to possess sufficient experience and undergo comprehensive training. Maintaining continuous awareness of the additional implants throughout the entire operation and prioritizing the preservation of soft tissues are of paramount significance. A profound comprehension of the system and its inherent constraints can also prove to be pivotal in certain situations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 2nd Department of Orthopaedics, "Hygeia" General Hospital of Athens, 15123 Athens, Greece.
- Department of Orthopaedics and Traumatology, "251" Hellenic Air Force General Hospital of Athens, Kanellopoulou Av 3, 11525 Athens, Greece.
- Department of Anatomy, School of Medicine, Faculty of Health Sciences, National and Kapodistrian University of Athens, 11527 Athens, Greece.
Zitieren
APA
Dretakis, K., & Koutserimpas, C. (2024). Pitfalls with the MAKO Robotic-Arm-Assisted Total Knee Arthroplasty. Medicina, 60(2), 262. https://doi.org/10.3390/medicina60020262
Vancouver
Dretakis K, Koutserimpas C. Pitfalls with the MAKO Robotic-Arm-Assisted Total Knee Arthroplasty. Medicina. 2024;60(2):262. doi:10.3390/medicina60020262
BibTeX
@article{Dretakis2024,
author = {Dretakis, Konstantinos and Koutserimpas, Christos},
title = {Pitfalls with the MAKO Robotic-Arm-Assisted Total Knee Arthroplasty},
journal = {Medicina},
year = {2024},
volume = {60},
number = {2},
pages = {262},
doi = {10.3390/medicina60020262}
}
EUVIMED