Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study
The International Journal of Medical Robotics and Computer Assisted Surgery · 2024-12 · 20 (6)
DOI: 10.1002/rcs.70029 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, online erschienen am 23.12.2024. Verfasst von Zakareya Gamie und 5 weiteren Autor:innen. Beteiligte Einrichtung: Tsiridis Orthopaedic Institute, ICAROS Clinic, Thessaloniki, Greece. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures, Femur, Osteoarthritis, Knee. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
We investigated the accuracy of targeted resection thickness in patients undergoing primary Total Knee Arthroplasty (TKA) using the ROSA robotic system.
Methods
Calliper measurements of the distal femur (DF), proximal tibia (PT), and posterior condyles (PC) were taken in 44 patients from June 2023 to January 2024.
Results
Planned and actual resection depth difference was 0.67 mm ± 0.6 mm (mean ± SD) ( p = 0.217) and 0.94 mm ± 1.15 mm ( p = 0.4) for medial and lateral DF, 0.93 mm ± 0.81 mm ( p = 0.001) and 0.89 mm ± 0.8 mm ( p = 0.008) for medial and lateral PT, and 1.1 mm ± 0.97 mm ( p = 0.001) and 1.04 mm ± 0.79 mm ( p = 0.001) for medial and lateral PC, respectively.
Conclusion
The ROSA robotic system can achieve a high degree of accuracy for planned resection thickness. Results are valid only for the imageless ROSA TKA in patients with primary knee osteoarthritis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Tsiridis Orthopaedic Institute, ICAROS Clinic, Thessaloniki, Greece.
- Centre of Orthopaedic and Regenerative Medicine (CORE), Center for Interdisciplinary Research and Innovation (CIRI), Aristotle University of Thessaloniki (AUTH), Balkan Center, Thessaloniki, Greece.
- Academic Orthopaedic Department, Aristotle University Medical School, General Hospital Papageorgiou, Thessaloniki, Greece.
- Nuffield Department of Orthopaedics, Rheumatology and Musculoskeletal Sciences (NDORMS), Oxford University Hospital NHS, Oxford, UK.
ORCID-Profile: Zakareya Gamie, Eustathios Kenanidis
Zitieren
APA
Gamie, Z., Kenanidis, E., Douvlis, G., Mylonakis, N., Maslaris, A., & Tsiridis, E. (2024). Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study. The International Journal of Medical Robotics and Computer Assisted Surgery, 20(6). https://doi.org/10.1002/rcs.70029
Vancouver
Gamie Z, Kenanidis E, Douvlis G, Mylonakis N, Maslaris A, Tsiridis E. Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study. The International Journal of Medical Robotics and Computer Assisted Surgery. 2024;20(6). doi:10.1002/rcs.70029
BibTeX
@article{Gamie2024,
author = {Gamie, Zakareya and Kenanidis, Eustathios and Douvlis, Georgios and Mylonakis, Nikolaos and Maslaris, Alexander and Tsiridis, Eleftherios},
title = {Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2024},
volume = {20},
number = {6},
doi = {10.1002/rcs.70029}
}
EUVIMED